43
8
Troubleshooting
CD/PMD Analyzer Source Error Messages
Warning/Error
Number
Description
Recommended Action
-11
Module reset error:
The nulling was not performed
correctly.
Restart your unit to solve the
problem.
-12
Wrong module ID:
The module returns the wrong ID.
Call EXFO for assistance.
-20
Module communication error:
Communication error with the
module.
Restart your unit to solve the
problem.
-25
Checksum error:
Checksum error while reading the
module’s FIFO.
Restart your unit to solve the
problem.
-30
Command not accepted:
The command that caused the
warning will be lost. The unit may
continue with the program even if the
command was not performed.
If problem persists, call EXFO for
assistance.
-31
Module setting error:
One of the settings sent to the unit is
wrong.
Review your command before
sending it again.
-32
Action currently in progress:
You cannot send a command while
the unit is already active.
Wait until the unit is done before
sending the command.
-34
Command overflow:
Too many commands were sent to
the unit at a time.
Wait until the unit is done before
sending more commands.