background image

 

 

AC8 & AC9 Series : Quick Start Guide

 

 

 

 

SIDE 1 

 

Warning – Refer to Safety Precautions Booklet 035-00636R for all safety precautions prior to use. 

User Guide 035-00563R download: 

http://www.excelitas.com/Pages/Product/OmniCure-AC-8-Series.aspx 

User Guide 035-00621R download: 

http://www.excelitas.com/Pages/Product/OmniCure-AC-9-Series.aspx

  

 

 

WARNING:

 Never look at operating LED!  The light could severely damage the cornea and retina of the eye if the light is observed directly. Eye shielding 

must be used at all times as well as clothing to protect exposed skin  

Printed in Canada 

Refer to user guide for more details. 

035-00614R Rev. 2 

 

SETUP / Unpack 

 

Carefully unpack the unit and accessories from shipping carton.   

 

When removing the LED Head, ensure that the optical window is not placed directly on 
a surface to avoid any damage

.

   

1.1

 

System Requirements 

In addition to the UV LED Head, you will require the following components: 

 

Power supply 

 

Power cable 

 

Note:

  You will also require a PLC controller unless you are using an existing controller. 

The PLC controller is used to discover the extensive functionality of the UV LED Head, 

but integrators can access the full feature set through integrating with the PLC 
interface directly. Refer to

 “Connecting the UV LED Head” 

documentation in the User 

Guide

 

Connecting the Power Supply and PLC Controller 

Note

:   All connectors are via screw style fastener -- do not over tighten the 

connector screws. 

Note

:   The following is a generic procedure, illustrated with the common 

components described above. Refer to

 

Connecting the Power Supply and 

PLC Controller” 

documentation in the User Guide 

Tip

:   After securing the power cable to the power supply, it is recommended to 

“tie-off” the female connector cable on to the mounting brackets or process 

rigging as a form of strain relief.  

1.

 

Make sure the power supply unit is turned off.  

2.

 

Connect the AC power cord to 

the power supply. 

3.

 

Attach the male end of the 

power cable to the power supply 
unit, and connect the female end 

of the power cable to the UV 
LED Head per 

Figure 1 – 

AC8300 

shown.  

 

 

UV LED Head Connection Ports 

 

1.2

 

Powering Up and Powering Down 

1.

 

Ensure that the system has been properly installed and the UV LED Head is securely 

mounted with the optical lens in the desired orientation. Refer to

 “Mounting the UV 

Head” 

documentation in the User Guide 

2.

 

Verify that the power supply unit’s AC supply cord is connected to a properly grounded 
AC outlet. 

3.

 

If the power supply has a circuit breaker, switch it to the “ON” position. 

4.

 

Turn the main power switch of the power supply to the “ON” position and check the fan 

for airflow. 

5.

 

The LED indicator on the UV LED Head will 

flash

 green during “boot-up” and change to a 

steady

 green indicating UV LED Head is ready for use. It is recommended to wait for 15 

seconds from initial power-up before attempting to control the UV head. For more 
information about the indicators. Refer to 

“Using the LED Head” 

in the User Guide 

1.3

 

Using the LED Head 

The peak wavelength of the UV LED Head is 365nm or 395nm +/- 5nm, depending on 
model. 

For all models, adjusting the UV LED Head intensity can be actuated by an external 
controller via the PLC interface. Refer to

 “Connecting UV LED Head” 

documentation in the User Guide 

Input voltage vs. desired intensity; 5V = 100% and 1V = 20%. A minimum voltage of 
1V is required for light source operation. 

Note:

 The light source will maintain a 20% intensity when the input voltage is in the 

range of 0.5V. If the input voltage is below this, the light source will turn off. 

Ensure the UV LED Head indicator is illuminated to a steady green before applying 
UV power to the head. The UV LED Head indicator will illuminate to a steady amber 

indicating UV is on. When UV is off, the indicator will revert back to a steady green 
indicating UV is off. 

The UV LED Head is capable of responding to PLC control signals in less than 0.2 
seconds.Understanding the LED Indicators 

PLC 