background image

 

 

 

X-Cite

®

 XLED1: Quick Start Guide

 

 

 

 

SIDE 1 

 

Warning 

– Refer to Safety Precautions Booklet 035-00636R for all safety precautions prior to use. 

User Guide 035-00465R download: 

http://www.excelitas.com/Pages/Product/X-Cite-XLED1.aspx

 

 

 

WARNING:

 Never look into the light emitting end of the light guide!  The light could severely damage the cornea and retina of the eye if the light is 

observed directly. Eye shielding must be used at all times as well as clothing to protect exposed skin  

Printed in Canada 

Refer to user guide for more details. 

035-00466R Rev. 4 

 

For Detailed Instructions refer to X-Cite

®

 XLED1 User Guide 035-00465R 

 

General Installation 

1. 

Install the LED Modules, Drivers and Dichroics, refer to User Guide 035-
00465R or 

http://www.excelitas.com/Pages/Product/X-Cite-LED-Module-

Configurator.aspx

 

2. 

Connect the XLED1 to your PC via the USB connection located on the rear 
panel and power on, allow the drivers to be installed from Windows 

Update. 

3. 

Install GUI Control Panel from the website, link provided in step 1, open 

the zip file and double click the setup.exe file 

4. 

If triggering is required, connect the optional 15pin HD-15 trigger I/O 
breakout cables to the trigger input and sync output connections located on 

the rear 

XLED1

 panel. 

5. 

Connect the optional Touch Screen Controller to the connector provided at 

the rear of the XLED1 panel. 

6. 

Connect the AC cable to AC connector at the rear 

XLED1

 panel. 

 

CAUTION  

To protect the LED Module, Driver and system from electrostatic discharge 

(ESD) related damage, which can disable the system, cause latent defects or 

permanent damage, or require repair by service technicians, please take the 

following ESD prevention steps:  

 

Ensure the unit is unplugged. 

 

Ensure you are properly grounded, ideally by using an ESD grounding strap 

attached to an ESD protective surface. 

 

LED Module/Driver and Dichroic Installation 

 

If modules need to be installed in the XLED1 System, please refer to User Guide  
035-00465R or 

http://www.excelitas.com/Pages/Product/X-Cite-LED-Module-

Configurator.aspx

 

 

1.

 

Ensure the system is OFF and the AC power cord is unplugged from the 

power source. 

2.

 

Using the locking hub key supplied, rotate the green locking hubs 90 

degrees and remove the locking hubs. Holding the XLED1 housing at the 

back, slide the XLED1 housing up and back (away from the fascia) for 

removal. The locking hub key can be stored on the locking hub key hanger, 

located on the back panel of XLED1. 

3.

 

LED Modules are shipped with their unique drivers. Both components must 

be installed at the same time. If there is a mismatch installation of either 

standard powered or high powered drivers, the corresponding LED will not 

operate. 

4.

 

Remove the protective LED lens cap. To install the LED Module and Driver, 

align the LED Module and Driver male pins with the female socket 

connection and gently push down into place. Tighten the Philips-Head 

screws until a positive force is felt and apply ¼ turn. 

DO NOT OVER 

TIGHTEN THE SCREWS

(If using a calibrated torque-meter, 

Excelitas Technologies recommends a torque value of 5.7 lb-in/± 

0.3 lb-in.)

 

5.

 

Install the Dichroics by inserting through the slot and lightly press down 

until a positive force is felt. Ensure to hand tighten the locking screw, do 

not use mechanical tooling such as screw drivers, pliers etc. 

6.

 

If the Dichroic is inserted incorrectly, the Dichroic’s orientation key (located 

on the frame), will not pass through the cut-out on the top plate, resulting 

in an unseated Dichroic. 

7.

 

When changing a Dichroic, the Edge value must be manually entered into 

the system (if using the optional Touch Screen Controller ). If using the 

GUI control panel Refer to User Guide 035-00465R for instructions on 

changing the edge, reflect and transmit values.  

8.

 

If using the optional Touch Screen Controller, to change the Edge value, 

touch on the Dichroic to see the pop-up window, touch the top tab of the 

window and an edit screen will appear. Type in the new Edge value and 
press “GO” to return to the maintenance screen. The Edge value is printed 

on the Dichroic module. For new Dichroics, the reflect and transmit values 

will not be automatically updated, refer to:  

 

http://www.excelitas.com/Pages/Product/X-Cite-LED-Module-

Configurator.aspx

 for these values. 

9.

 

Replace the XLED1 housing ensuring a snug fit. Insert the locking hubs and 

using locking hub key supplied, rotate the green locking hubs to the locked 

position. Return the locking hub key to the locking hub key hanger, located 

at the rear of the system. The system will not allow the main power switch 

to function if the housing is not replaced properly. 

 

Installing the Light Guide 

1.

 

Ensure the Power is not connected to the 

XLED1

2.

 

Remove the protective end caps from the Light Guide, both input and 

output ends. 

3.

 

Holding the Light Guide by the strain relief portion only, insert the Light 

Guide into the light guide port on the front panel until it seats with a 

positive click (up to the “A” line printed on the grey ferrule).  

4.

 

When the Light Guide is fully inserted and when the power is ON, the 

Light 

Guide LED

 will illuminate 

green

 indicating full Light Guide insertion. If the 

Light Guide is not fully inserted the light guide indicator LED will be RED. 

5.

 

To remove the Light Guide grip the strain relief portion only and pull out 

firmly. 

 

 

 