MFC
Maintenance
7.
207
Replacing electrolyte,
electrodes and diaphragms
1
Remove DEPOLOX
®
5 flow block assembly cover.
2
Close the ball valve on the sample water inlet and on the outlet
(pressurised version).
3
Open the drain valve and drain the cell body (hold container
underneath).
4
Close the drain valve when the cell body is empty.
5
Remove the sensors. Loosen the cable union (hold the cable
while doing this as it may not be allowed to rotate).
6
Loosen the lower cap on the 3 electrode cell.
7
Remove the signal cable.
Replace electrolyte
8
Unscrew the upper knurled nut on the electrolyte container.
9
Remove the electrolyte tank out of the cell body from below
using the electrode mount.
Please note
Wash the cell sand out of the electrode mount “Replacing cell sand
in DEPOLOX® 5 three-electrode cell" on page 205 “.
10
Remove the electrolyte container from the electrode mount.
To do this, unscrew the knurled nut in the electrode mount.
11
Remove the electrolyte container out the electrode mount from
the top.
12
Remove the drain plug from the electrolyte container.
13
Turn the electrolyte container upside down and drainthe KCI
electrolytes by lightly shaking it.
Replacing reference electrode
14
If necessary, replace the reference electrode.
The reference electrode can be unscrewed from the
electrolyte container. Lightly wet the O-ring before installinga
new reference electrode.
Replacing diaphragms
Please note
The diaphragms, which form the contact between the reference
electrodes and sample water, cannot be cleaned. If the water
quality is very good, the diaphragms can remain installed for up to
three years; they should be replaced thereafter (no exceptions). If
the sample water quality is poor, the diaphragms may be soiled.
This influences the measuring accuracy.
15
Remove both diaphragms from the electrolyte container using
a suitable tool (e.g. tweezers).
16
Push new diaphragms into the electrolyte container. Lightly
wet the O-rings.
17
Insert the electrolyte container back into the electrode mount.
Lightly wet the O-ring here as well.
Содержание WALLACE & TIERNAN MFC
Страница 1: ...WALLACE TIERNAN MULTI FUNCTION CONTROLLER MFC from version V 2 00 INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 580 000 DE IM 0714 MFC Please note Original manual...
Страница 6: ...6 WT 050 580 000 DE IM 0714 Introduction MFC 1...
Страница 40: ...40 WT 050 580 000 DE IM 0714 Description MFC 3...
Страница 225: ...MFC Complete Devices Retrofit Kits and Spare Parts 8 225 DEPOLOX 5 drawing non pressurised version W3T170031 W3T170031...
Страница 229: ...MFC Complete Devices Retrofit Kits and Spare Parts 8 229 DEPOLOX 5 drawing pressurised version W3T170032 W3T170032...
Страница 233: ...MFC Complete Devices Retrofit Kits and Spare Parts 8 233 VariaSens drawing non pressurised version W3T158742 W3T158742...
Страница 237: ...MFC Complete Devices Retrofit Kits and Spare Parts 8 237 VariaSens drawing pressurised version W3T158756 W3T158756...
Страница 247: ...MFC Certificates 10 247 10 Certificates...
Страница 248: ...248 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Страница 249: ...MFC Certificates 10 249...
Страница 250: ...250 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Страница 262: ...262 WT 050 580 000 DE IM 0714 Settings table MFC 11...
Страница 266: ...266 WT 050 580 000 DE IM 0714 Index MFC 12...