DEPOLOX
®
5 E flow cell
Installation
4.
41
4.4.10
Removing the felt ring
To keep the diaphragm moist and prevent crystallization in the electrolyte
solution there is a moist felt washer in the gap between electrolyte tank
and electrode when the unit is in storage.
Attention!
Before initial startup, the felt washer for moistening the membranes must
be removed.
Proceed as follows:
1
Extract the knurled nut (A) from the cover of the electrolyte tank.
2
Remove the entire electrode unit from the cell body from below.
3
Remove the felt washer (B) between electrolyte tank and electrode.
4
Refit the entire electrode unit in the cell body. With the locating pin,
observe the exact position in the cell body.
5
Push the entire electrode unit upwards.
6
Refit the knurled nut on the electrolyte tank.
Image 7 Cross-section of cell body, non-pressurized version (example
knurled nut)
A Knurled nut
B Felt washer
A
B