background image

Halocarbon  Refrigerants  Detector-TransmitterE2611-HFC

  is  intended  for  early 

detection of leakages or accumulation of halocarbon refrigerants in indoor spaces. 
The  instrument  is  based  on  fully  calibrated  and  temperature  compensated 
semiconductor metal oxide sensor with high repeatability, stability and long lifetime.  
E2611 provides two independent analog outputs OUT1 and OUT2, user-selectable 
to  4-20  mA  or  0-10  V,  proportional  either  to  gas  concentration  or  temperature. 
RS485  Modbus  RTU  digital  communication  interface  allows  easy  instrument 
configuration and integration into various automation systems. Two relays RE1 and 
RE2  with  closing  contacts can  be  used  to  switch  24  V  or  230  V  powered  alarm 
sirens, ventilation fans, shut-off valves or other actuators. 

Safety requirements 

Always adhere to the safety provisions applicable in the country of use. 
Do not perform any maintenance operation with the power on. Do not let water or 
foreign objects inside the device. 

Operating conditions 

 

The device should be used in explosion-safe (non ATEX -rated) indoor areas at the 
atmospheric  pressure  ±10%,  <95  %RH,  without  condensation,  and  temperature 
within the range -30...+70 °C. 

 

Metal  oxide  sensor  cannot  properly  operate  in  a  zero  or  low  oxygen  content 
atmosphere. Normal ambient oxygen concentration is recommended.  

 

Avoid  exposure  to  highly  corrosive  gases  (H

2

S,  SO

2

,  HCl,  Cl

2

  etc),  and  volatile 

silicon  containing  materials  (silicone  rubber/putty,  hair  grooming  materials, 
adhesives). 

 

Avoid mechanical shock or strong vibrations. 

 

Avoid sources of electromagnetic interference 

Installation and connections 

The gas detector should be mounted on a wall at a place located not more than 5 
m  from  a  possible  leakage  source  and  not  very  close  to  ventilation  openings  or 
strong air currents. Avoid the areas without air circulation (corners, niches) as well. 
Since  halocarbon  refrigerants  are  heavier  than  air,  it  is  advisable  to  place  the 
detector  not  higher  than  a  potential  leakage  source.  Consider,  if  the  vapors  are 
heated or not, if the detector is used for fire safety (LEL range) or air quality control 
(ppm range) etc. Take into consideration the geometry of the room and the speed 
and direction of air currents.  
Make sure that the installed detector is available for maintenance and repair. 
To install and connect the device, proceed as follows: 
1. Detach the base of the enclosure by gently pulling along four guiding pins. 
2. Attach the base to the wall with two screws. (see dimensional drawing below). 

 

3. Connect power supply and external devices to the terminal blocks on the PCB 

according to the connection diagram below.  

 

To power the instrument from an external 24 VDC source, connect terminals 0 V 
and +U to the source. If an integrated mains power supply module is used, connect 
terminals L and N to the mains. 

NB!

 If the instrument is powered from mains, connect to 0 V and +U terminals only 

light  external  loads,  which  consume  less  than  30  mA  in  total,  as  the  integrated 

mains supply module capacity is limited. 
To use analog outputs, connect the terminals OUT1 and/or OUT2 and 0 V to the 
input of the secondary instrument (indicator or controller). 

NB!

 The outputs are not  galvanically isolated from 24 V power supply and share 

common 0 V. Allowed load resistance limits are stated in Specifications table. 
The type of each analog output can be independently changed between 4-20 mA 
and 0-10 V with jumpers J1 (OUT1) and J2 (OUT2). 
With closed jumper the output is 0-10 V, with open jumper the output is 4-20 mA. 
By default, both outputs OUT1 and OUT2 are assigned to gas concentration. The 
device has built-in temperature sensor which may be tied to any of the outputs. The 
output assignments and scales can be changed by Modbus commands.  
We  recommend  to  set  the  difference  between  the  upper  and  bottom  limits  of the 
output scale not narrower than 20% of detection range. In any case, do not set the 
output scale below the tenfold resolution of the device. 
To  use  relay  outputs,  connect  the  chosen  actuators  to  the  relay  terminals  RE1 
and/or RE2.  

NB! 

Actuator  short-circuits shall  be  avoided,  to  protect the  instrument  relays  use 

external fuses or safety switches. 
4.  Turn  on  the  power.  It  may  take  up  to  five  minutes  after  switching  on  for  the 
sensor to stabilize. During this period relays, analog outputs and Modbus interface 
are  off. A  LED  placed  on  the  PCB  of  the  device  allows to  control  the connection 
process. The LED response to different processes is presented in the table below. 

Process 

LED mode 

Sensor heating period 

Blinking 0.5 Hz (50% on, 50% off) 

Sensor absence or malfunction 

Blinking 0.5Hz (90% off, 10% on) 

Relay 1 turned on 

Blinking 1 Hz (50% on, 50% off) 

Relay 2 turned on 

Blinking 2 Hz (50% on, 50% off) 

Modbus response 

The signal is modulated with short on off 
pulses, even single Modbus cycle is 
traceable* 

Normal measurement 

Continuous light 

* Pulse corresponding to Modbus response is visible only when diode light is on 

5. Push enclosure to the base. 
 

 

Operation 

For best stability the gas detector should be powered permanently. If the instrument 
is left for a long time in unpowered state, then after initial power-on the metal-oxide 
gas sensor needs some time to heat up and burn adsorbed contaminants. So for 
first tens of seconds of few minutes an alarm activation may be possible. After this 
heating-up period the instrument turns into normal mode. 
When  the  concentration  of  the  detected  gas  reaches  the  LOW  alarm  level,  the 
control  LED  starts  blinking  and  the  buzzer  starts  beeping  with  1  Hz  frequency. 
When the HI alarm level is reached, the frequency of blinking/beeping is 2 Hz. The 
alarm signal turns off automatically, if gas concentration decreases to 80% of the 
alarm set point. 

Configuring 

Gas  detector  E2611  shares  all  functionalities  of  the  PluraSens®  multifunctional 
transmitter platform. The features and options include: 
- digital output change rate limiting filter 
- digital integrating (averaging) filter 
- temperature measurement channel with internal sensor 
- free assignment of each analog output to chosen parameter 
- flexible setting of analog output scales for each output 
- output shift and slope adjustment for calibration 
- free assignment of each of two relays to chosen parameter 
- several relay control logic modes (HI or LO with hysteresis, U or П) 
- switch delays and minimum on/off state durations for each relay 
- Modbus controlled forced state option for analog outputs and relays. 
E2611 can be configured through its RS485 interface by Modbus RTU commands. 
A  standard  configuration  kit  includes  Model  E1087  USB-RS485  converter  and 
E26XX Configurator software.  
For procedure details please contact your Seller or Manufacturer. 

Calibration and maintenance 

The device has been calibrated by Manufacturer with standard gas mixtures before 
delivery. The instrument requires at least annual field recalibration with a portable 
calibration  kit.  For  procedure  details  please  contact  your  Seller  or  Manufacturer. 
The  calibration  should  be  preferably  performed  by  Manufacturer's  authorized 
representative. 

Return to default settings 

To  reset  the  device’s  Slave  ID,  baud  rate  and  sbit  number  to  factory  settings, 
proceed as follows: 
1. De-energize the device 
2. Connect the J3 jumper 
3. Turn on the device 
4. De-energize the device 
5. Disconnect the J3 jumper 
6. Turn on the device 

Warranty 

This product is warranted to be free from defects in material and workmanship for a 
period  of  one  year  from  the  date  of  original  sale.  During  this  warranty  period 
Manufacturer  will,  at  its  option,  either  repair  or  replace  product that  proves  to  be 
defective.  This  warranty  is  void  if  the  product  has  been  operated  in  conditions 
outside  ranges  specified  by  Manufacturer  or  damaged  by  customer  error  or 
negligence or if there has been an unauthorized modification.

 

 

85

37

8

5

63

Input 1 4-20 mA / 0-10 V

Input 2 4-20 mA / 0-10 V

0V / GND

Controller

OUT1
OUT2
0V

A
B

RS485

+U = 11...30 VDC

L
N

90...265 VAC

RE1

RE2

Fieldbus

S

PSU

24 VDC

+

-

J1 J2

230 VAC

J3

Control LED

Отзывы: