Specifications
Sampling method
Diffusion
Sensor type
Standard: metal oxide semiconductor
Typical detection range
0...1000 ppm
Resolution / digital unit
1 ppm
Response time T90
<120 s
Sensor lifetime
> 5 years
Calibration interval
12 months
Signal update
Every 1 second
Power supply options
11...30 VDC (default),
24 VAC or 90...265 VAC as options
Power consumption
< 2 VA
Analog outputs
2 × 4-20 mA / 0-10 V, user settable
Outputs assignment
OUT1: Gas; OUT2: Gas
Load resistance
R
L
< (Us - 2 V) / 22 mA for 4-20 mA
R
L
> 250 kOhm for 0-10 V mode
Digital interface
RS485, Modbus RTU protocol
No galvanic isolation
Output scale width
> 10 × resolution
Recommended: 20-100% of the detection range
Relay outputs
2 × SPST, max 5 A, 30 VDC / 250 VAC
Default alarm setpoints
RE1 (LOW): set 100 ppm; release 80 ppm
RE2 (HIGH): set 500 ppm; release 400 ppm
Enclosure
Light beige ABS, wall mount, protection class IP65
Dimensions
H87 × W82 × D55 mm
CE marking
According to 2014/30/EU and 2014/35/EU, EN 50491-4-1:2012
EN 61000-6-3:2020, EN 61326-1:2013(EMC, emissions)
EN 61000-6-1:2019, EN 61000-6-2:2019(EMC, Immunity)
EN 60079-29-1:2016, EN 60079-29-2:2015 and EN 60079-29-3:2014
Operating conditions
-30...+60 °C; <95% RH non-condensing, 0,9...1,1 atm
Explosion- safe indoor areas,
Normal ambient oxygen level
Avoid strong mechanical shock, vibrations or EMI
Avoid exposure to corrosive gases or silicon containing products
5