MULTIPOINT CIRCULAR-CHART RECORDER: USER GUIDE
HA250385
Issue 7 Nov 06
Page 94
9.4 SPECIFICATION
Inputs
Range
±
100 mV and 0 - 10 V dc (auto ranging)
Sample rate
9 Hz (110 ms)
Calibration accuracy
0.25% of reading,
±
1 LSD,
±
1˚C/F
Resolution
<1
µ
V for 100 mV range, <0.2 mV for 10 V dc range
Linearisation accuracy
<0.1% of reading
Input filter
1.0 to 999.9 seconds
Zero offset
User adjustable over the full display range
Thermocouple
Types:
J, K, T, L, N, C, R, S, B & Platinell II
Cold junction comp:
Automatic: typically >30 to 1 rejection from 20 ˚C
RTD
Types:
Platinum 100, DIN std. 0.00385
Ω
/
Ω
/˚C
Excitation current:
0.2 mA
Lead compensation:
Up to 22
Ω
in each of 3 equal resistance leads
V, mV ranges
±
100 mV and 0 - 10 V dc (user configurable with these limits)
Outputs
Relay
Rating:
Min: 12V @ 100 mA, Max: 2 A @ 264 Vac resistive
Triac
Rating:
1 A resistive @ 264 Vac
Analog
Range:
Isolated, 0 to 20 mA configurable (load 600
Ω
max.), 12 V dc limit
Control Functions
Control
Modes:
PID or PI with overshoot inhibition, PD, PI, P only or On/Off
Auto/manual:
Bumpless transfer
Setpoint rate limit:
0.01 to 99.99 degrees (or display units per minute)
Cooling algorithms:
Linear, Water (non-linear), Fan (minimum on time), Oil, proportional only
Tuning
One-shot tune:
Automatic calculation of PID and overshoot inhibition parameters
Automatic droop comp.:
Automatic calculation of manual reset value when using PD control
Alarms
Types:
Full scale high or low, Deviation high, low or band
Modes:
Latching or non-latching, Normal or blocking action
up to 4 process alarms can be combined onto a single output
General
Display
Dual, 4-digit x 7-segment high intensity LED
Dimensions & weight
48 mm W x 48 mm H x 103 mm D (1.89" x 1.89" x 4.06"); 250 g (8.82 oz)
Supply
100 to 240 Vac (-15%,+10%), 48 to 62 Hz, 10 Watts max.
Environmental
Temperature:
Operating: 0 to 40 ˚C (32 to 104 ˚F) (ambient with controller); Storage -10 to 70 ˚C
Humidity:
5 to 90% RH (non condensing)
Atmosphere:
Electrically conductive pollution must be excluded from the controller cabinet
Altitude:
Not suitable for use above 2000 m (6,562 ft) or in corrosive or explosive areas
Panel sealing
IP65, NEMA 4X
Electromagnetic compatibility
BS EN50081 general emissions for industrial environments
BS EN50082-2(95) standards for industrial environments
Safety standards
BS EN61010, installation category 2 (voltage transients not to exceed 2.5 kV)
Содержание 394
Страница 1: ...Model 394 Multi point circular chart recorder User Guide EUROT HERM ...
Страница 4: ......
Страница 122: ...MULTIPOINT CIRCULAR CHART RECORDER USER GUIDE HA250385 Issue 7 Nov 06 Page 118 This page is deliberately left blank ...
Страница 138: ...HA250385 Issue 7 Nov 06 Page 134 MULTIPOINT CIRCULAR CHART RECORDER USER GUIDE This page is deliberately left blank ...
Страница 140: ......
Страница 141: ......
Страница 142: ......