6 Operation
50/84
Instructions for use
•
MERGITE!
WARNING – PRESSURE COMPENSATION HOLE
The caps of the containers have a pressure compensation hole.
Excess pressure in the waste liquid container may impair the
discharge of the waste liquid.
>
Do not seal the pressure compensation hole.
1) Place a paper towel at the ready to mop up any dripping liquid
residues.
2) Disconnect the full container.
3) Dispose of the contents of the full container in accordance with
the statutory regulations.
4) Reconnect the drained container.
5) If you were prompted by the system to drain the waste liquid
container, finish the process by pressing
OK
.
The waste liquid container has been drained.
Filling the container for distilled water
Filling the container for distilled water is necessary when the fill level
of the container has reached its lower limit.
1) Place a paper towel at the ready to mop up any dripping liquid
residues.
2) Disconnect the empty container.
3) Dispose of the residue content of the container.
4) Fill the container with distilled water.
5) Reconnect the filled container.
6) If you were prompted by the system to replace the container for
distilled water, finish the process with
OK
.
You have filled the container for distilled water.
Filling the wash buffer container
Filling the wash buffer container is necessary when the fill level of the
container has reached its lower limit.