background image

 

 

Model HI-1801 

Microwave Survey Meter 

User Manual 

 

 

 

 

 

 

 

Содержание HI-1801

Страница 1: ...Model HI 1801 Microwave Survey Meter User Manual ...

Страница 2: ...rs Copyright 1990 2016 by ETS Lindgren L P All Rights Reserved No part of this document may be copied by any means without written permission from ETS Lindgren L P Trademarks used in this document The ETS Lindgren logo is a trademark of ETS Lindgren L P Revision Record HI 1801 Microwave Survey Meter MANUAL Part H 600041 Rev C Revision Description Date Initial Release February 1990 A Revised Reform...

Страница 3: ...ions 9 Battery Replacement 9 Spacer Cone Replacement 10 Annual Calibration 10 Replacement and Optional Parts 11 Service Procedures 11 3 0 Specifications 13 Electrical Specifications 13 Physical Specifications 13 4 0 Operation 15 Accuracy 15 Preparing the Oven 17 Preparing the Meter 18 Check Feature 19 Making a Measurement 19 What the Measurement Means 20 Appendix A Warranty 21 Appendix B EC Declar...

Страница 4: ...iv This page intentionally left blank ...

Страница 5: ...ons could result in minor personal injury and or property damage Included text gives proper procedures Warning Denotes a hazard Failure to follow instructions could result in SEVERE personal injury and or property damage Included text gives proper procedures Note See the ETS Lindgren Product Information Bulletin for safety regulatory and other product marking information ...

Страница 6: ...vi This page intentionally left blank ...

Страница 7: ...ular to the axis of the probe The antenna lobe effective measuring area is also very broad making the instrument easy to use when measuring leakage around an oven door The spacer cone is designed to provide 5 cm spacing from the tip of the probe to the center of the array The HI 1801 must be used with the spacer cone in place Using the instrument without the cone will result in reading errors and ...

Страница 8: ...ing case Optional Items Spacer Cover Kit ETS Lindgren Product Information Bulletin See the ETS Lindgren Product Information Bulletin included with your shipment for the following Warranty information Safety regulatory and other product marking information Steps to receive your shipment Steps to return a component for service ETS Lindgren calibration service ETS Lindgren contact information ...

Страница 9: ...rvey Meter is limited to external components such as cables or connectors If you have any questions concerning maintenance contact ETS Lindgren Customer Service Maintenance Recommendations The only maintenance that is required is the replacement of the batteries or the spacer cone BATTERY REPLACEMENT Always replace both batteries when batteries are in need of replacement 1 Move the selector switch...

Страница 10: ...pacer cone in position To maintain stated instrument accuracy only use spacer cones provided by ETS Lindgren The spacer cone must be replaced when worn or contaminated A polystyrene cone shield provides protection for the relatively soft material of the cone 1 Remove the old cone by pulling straight up and off the end of the probe 2 Slip a new cover on being sure the cover is fully seated on the p...

Страница 11: ...ng parts Following are the part numbers for ordering replacement or optional parts for the model HI 1801 Microwave Survey Meter Part Description Part Number Spacer Cover Kit 10 Cones and 20 Shields H 540013 600 mL Plastic Beaker H 44600MLP Carrying Case H 51FPC6506B Service Procedures For the steps to return a system or system component to ETS Lindgren for service see the Product Information Bulle...

Страница 12: ...12 Maintenance This page intentionally left blank ...

Страница 13: ... Range 1 10 mW cm2 Accuracy 1 dB Response Time Full scale within 3 seconds Overload Capacity Continuous 2 0 W cm2 Physical Specifications Dimensions 5 3 x 4 3 x 10 5 cm 2 09 x 1 7 x 413 in Probe Length 30 5 cm 12 in Cable Length 1 0 m 3 3 ft Spacer Cover 5cm 1 97 in Shipping Weight 1 81 kg 4 lbs ...

Страница 14: ...14 Specifications This page intentionally left blank ...

Страница 15: ...mounting gasket may cause leakage in the control panel area of the oven Accuracy The accuracy of the ETS Lindgren Model HI 1801 Microwave Survey Meter is within 1 db 25 20 when used according to these instructions 1 Always use a clean spacer cover Accuracy will be affected by wear of the cover and by dirt and metallic particles which may become imbedded in the EPS expanded polystyrene foam If the ...

Страница 16: ... reads high at lower temperatures and low at higher temperatures 4 The parameters which affect the accuracy of this instrument are listed along with the error contribution of each in the following table Parameter Error dB Calibration Precision 0 09 Accuracy 0 13 Nonlinearity and AM response 0 17 Near Field vs Far Field 0 29 Receiving Pattern 0 11 Temperature Response 0 06 Frequency Response 0 04 P...

Страница 17: ... oven is set for full power 4 In order to make an accurate leakage measurement you must use the specified water load Fill the beaker to a level of 275 mL with cool tap water and place it in the center of the oven 5 Set the oven timer for approximately three minutes If your test takes longer than this the water may boil If the water boils pour it out and refill using 275 mL of cool tap water If the...

Страница 18: ...come on when the ON OFF button is held in either position and the instrument is operating 3 Turn the ON OFF switch to either the ON or the CHECK position The LOW BATT light will come on if the batteries need to be replaced Replace the batteries before continuing See the Maintenance section for battery replacement instructions 4 While operating the ON OFF switch in the ON 10 MW position use the lar...

Страница 19: ...2 Holding the ON OFF switch in the CHECK position and the oven operating move the probe over the door and door gap surfaces of the oven In the CHECK mode the sensitivity of the HI 1801 is increased from five to ten fold Making a Measurement Turn the oven on and proceed with the leakage measurement 1 Hold the probe by the red handle in one hand and hold the meter in the other hand with the ON OFF s...

Страница 20: ...5 mW cm2 measurement is mid scale on the 10 mW range However many ovens leak so little that you may notice only a small indication This actual value is sometimes as low as 0 1 or 0 2 mW cm2 If the oven leakage exceeds 5 mW cm2 you should contact a service agency recommended by the oven manufacturer or contact the manufacturer directly This oven must be repaired to comply with the CDRH safety guide...

Страница 21: ...TS Lindgren warranty for your HI 1801 Microwave Survey Meter DURATION OF WARRANTIES FOR HI 1801 MICROWAVE SURVEY METER All product warranties except the warranty of title and all remedies for warranty failures are limited to one year Product Warranted Duration of Warranty Period HI 1801 Microwave Survey Meter 1 Year ...

Страница 22: ...22 Warranty This page intentionally left blank ...

Страница 23: ...EC Declaration of Conformity 23 Appendix B EC Declaration of Conformity ...

Страница 24: ...24 EC Declaration of Conformity This page intentionally left blank ...