INSTRUMENT OPERATION

 

- Simply aim the RayTemp 6 Infrared (IR) thermometer at 

the target and press the trigger to measure the surface temperature. When the trigger is 

released the reading is automatically held for 60 seconds and 'HOLD' will be displayed, after 

which the thermometer will shut down. 

MEASUREMENT ZONE/TARGET DISTANCE

 - The measurement zone is proportional to the 

distance the RayTemp is away from the target. The RayTemp 6 is equipped with a 12:1 lens. If 

the target is 60 cm (24") away the measurement zone will be 5 cm (2") across.

EMISSIVITY

 - The RayTemp 6 has a default emissivity of 0.95 but is adjustable from 0.10 to 

1.00.  The closer the emissivity value is set to the actual emissivity value of the object being 

measured, the more accurate the readings will be. To adjust the emissivity press the MODE 

button so that 'E' is displayed, press the button again so that '

' is displayed. Use the UP   

and DOWN   buttons to set the emissivity required pressing the MODE button to set.

Please note:

 non-contact infrared thermometers are not recommended for use in 

measuring shiny or polished surfaces.

LASER POINTER

 - The circular laser function can be turned on or off by pressing the trigger 

and simultaneously pressing the °C/°F button whilst the trigger is being pressed. The laser 

icon '

' will be shown to indicate on or off. The centre of the measurement zone is 14 mm  

(approximately ½ ") above the lowest laser dot with the instrument held in a horizontal 

position. For safety, the laser pointer will only activate when the trigger button is pressed. 

The laser module is a Class 3 device that has a maximum power output of less than 5 mW at 

a wavelength of 660 nm. Prolonged, continuous exposure, such as staring at the beam, can 

be harmful and should be avoided. DO NOT look at the beam of any optical instrument.

LCD BACKLIGHT

 - The LCD backlight function can be turned on or off by pressing the 

trigger and simultaneously pressing the LOCK button. The backlight icon '

' will be shown 

to indicate on or off.

UNDER RANGE/OVER RANGE 

- If the measured temperature goes below or above the 

range of the instrument, 'Er' will be displayed. When the measured temperature returns 

within the range of the instrument, readings will automatically be displayed.

MAXIMUM, MINIMUM, DIFFERENTIAL & AVERAGE

 - To access the maximum reading, press 

the MODE button until 'MAX' is displayed. To access the minimum reading, press the MODE 

button again until 'MIN' is displayed. To access the differential reading, press the MODE 

button until 'dIF' is displayed. To access the average reading, press the MODE button again 

until 'AVG' is displayed.  Maximum and minimum, differential and average readings will be 

reset when the instrument is turned off.

°C/°F

 - Press the °C/°F button without pressing the trigger until the °C or °F scale changes.  

Please note:

 when the unit is next turned on, it will default to the last selected scale.

814-075/24.10.18