INSTRUMENT OPERATION

 - Simply aim the thermometer at the target and press the 

'Measure' button to display the surface temperature. The distance to target ratio is 1:1 
therefore the thermometer should be positioned as close to the target as possible.

Please note:

 Prolonged use at close range to a source of high or low temperature may 

affect the accuracy of the instrument due to thermal radiation.

MIN/MAX 

- First turn the instrument on by pressing the 'Measure' button, then press the 

'Mode' button once for maximum or twice for minimum. The 'MAX' or 'MIN' icon will 
flash. Press and hold the 'Measure' button; the thermometer will display the minimum or 
maximum reading only.

LOCK

 - The lock mode is particularly useful for continuous monitoring of temperatures. 

To utilise the lock mode, first turn the instrument on by pressing the 'Measure' button, 
then press the 'Mode' button three times for the lock mode function. The 'LOCK' icon 
will flash, then press the 'Measure' button to confirm the lock measurement mode. The 
thermometer will continuously display the temperature for up to 60 minutes or until the 
'Measure' button is pressed. Press the mode button to return to standard measurement.

°C/°F 

- To change the thermometer from '°C' to '°F' or from '°F' to '°C', first turn the 

instrument on by pressing the 'Measure' button, then press the 'Mode' button four times, 
the '°C' or '°F' symbol will flash, press the 'Measure' button to change the scale.

EMISSIVITY 

- The IR-Pocket thermometer is supplied with a default emissivity of 0.95. 

The emissivity of the thermometer can be changed from 0.1 (10E) to 1 (100E). Changes 
should only be carried out by experienced personnel. To change the emissivity, first turn 
the instrument on by pressing the 'Measure' button, then press the 'Mode' button five times 
for the emissivity function. '95E' will flash on the LCD screen, press the 'Measure' button 
to adjust the emissivity value, then press the 'Mode' button again to exit the set up screen. 
For information relating to the emissivity of specific materials, please visit our website 
or contact the ETI Sales Office. Please note: non-contact infrared thermometers are not 
recommended for use in measuring the temperature of shiny or polished metals.

LED SPOT ALIGNMENT 

- The LED indicator is not exactly the same as the target size. To 

get the best measurement result, aim at the target within a distance of about 2.5 cm.

IR-Pocket
Infrared Thermometer

814-060/31.05.16

www.etiltd.com

Operating Instructions

order code: 814-060