ETAS
Technical Data
ES441.1 - User Guide
100
8.10.3
Sensor Power Supply
Features
Characteristics
Sensor power supply channels
4, separate per input channel
Output voltage
Each channel can be set separately:
“Off”, in steps (5 V, 8 V, 10 V, 12 V,
15 V) or freely selectable as a value in
the range 5 V to 15 V
D/A converter
16-bit (resolution 229
V)
Protection
Outputs (plus) short-circuit proof
against ground and up to 26 V over
-
voltage.
Ground connected to operating volt
-
age ground and protected by a fuse.
Diagnostics
Short-circuit detection per channel;
overload conditions reduce the output
voltage
Parameter
Min
Max
Unit
Accuracy of the sensor supply voltage
without load
-
±10
mV
Accuracy of the sensor supply voltage
with 3 k
load
-
±5
mV
Resistance of sensor supply voltage
outputs including CBAV411.1-2 cable
-
1.6
Ohm
Ripple voltage V
pp
(measurement lim
-
ited to 80 MHz)
-
20
mV
Max. voltage drift (temperature)
-1
+1.2
mV/K
Output current
-
30
mA
Output current protection
-
200
mA