Checks and Maintenance
82
Platinous User’s Manual
7.4 Consumable Parts and Replacement Interval
The below parts must be replaced periodically. Promptly replace them at the specified intervals if not
sooner. Maintenance service is also provided by Espec North America. For spare parts and servicing,
contact the place of purchase or
.
TABLE 7.16.
C
ONSUMABLE
PARTS
AND
REPLACEMENT
INTERVAL
Part
Replacement Interval
Replacement Procedure
Wet-Bulb Wick
After temperature tests or every
month
See “Test Preparation” on
page 54.
Door Gasket (inner/outer)
3 years
Contact the place of purchase or
ESPEC.
Chamber Lamp
Every 6,000 hours or when
blown
See “Operator Error and Addi-
tional Troubleshooting” on
page 95.