Platinous User’s Manual
59
Operation
6.2 Specimen Safety Device Setup
6.2.1 Overheat Protector
Always set the overheat protector BEFORE beginning tests. Set it
according to the type of specimens and test conditions.
Unless properly set, the overheat protector will not trip in the event of
chamber trouble, which could lead to specimen damage. For details on
the overheat protector setting,
See “Specimen Protection (Safety
Procedure:
1. Turn the dials until a suitable temperature is displayed.
FIGURE 6.38.
O
VERHEAT
PROTECTOR
6.2.2 Absolute High/Low Limit and Upper Deviation Limit Temperature
(& Humidity) Alarms
Always set the absolute high/low limit temperature (& humidity)
alarms BEFORE beginning tests. Set alarms according to the type
of specimens and test conditions.
Unless properly set, the absolute high/low limit temperature (humidity)
alarms will not be triggered in the event of chamber trouble, which could
lead to specimen damage. For details on alarm settings,
Protection (Safety Devices)” on page 8.
!
CAUTION
Dial
!
CAUTION
Содержание Platinous ESL Series
Страница 1: ...PLATINOUS CHAMBER SERIES USER S MANUAL...
Страница 2: ......
Страница 6: ...iv Platinous User s Manual...
Страница 22: ...Precautions in Usage 16 Platinous User s Manual...
Страница 31: ...Platinous User s Manual 25 Names and Functions of Parts...
Страница 40: ...Names and Functions of Parts 34 Platinous User s Manual...
Страница 58: ...Installation 52 Platinous User s Manual...
Страница 80: ...Operation 74 Platinous User s Manual...
Страница 106: ...Troubleshooting 100 Platinous User s Manual...
Страница 107: ...Platinous User s Manual Chapter 9 Appendix...