Platinous User’s Manual
57
Operation
FIGURE 6.35.
H
OW
TO
FEED
SPECIMEN
CABLES
THROUGH
THE
CABLE
PORT
Use a specimen cable that can withstand test conditions. Refer to the following table for possible cable
types.
TABLE 6.12.
C
ABLE
TYPE
AND
WITHSTAND
TEMPERATURE
When not using the cable port, fit a silicone rubber plug in it from the inside and outside.
FIGURE 6.36.
W
HEN
NOT
USING
CABLE
PORTS
6.1.3 Wet-Bulb Wick Check (ESL, ESX models only)
The wet-bulb wick (included) must be installed to run humidity tests. If already in place, check whether it is
wet or dry before starting. If dry, change it.
Note:
Bacteria adhering to the wet-bulb wick can proliferate during tests and block water
supply. Wash hands with soap and water before handling the wick.
Procedure:
1. Remove the wet-bulb wick from its package.
2. Slide the wet-bulb wick over the wet-bulb temperature sensor. It is easier to slide the wick over the
sensor by folding it in half.
Specimen
Rubber
plug
Specimen
cable
To
specimen
power
supply
control
terminal
and
power
supply
Cable Type
Withstand
Temperature
Vinyl Cable
55°C
Styrene-butadiene cabtyre cable
70°C
Electron crosslinked heat-proof plastic cable
100°C
Glass-braid silicon shielded cable
200°C
Rubber
plugs
Содержание Platinous ESL Series
Страница 1: ...PLATINOUS CHAMBER SERIES USER S MANUAL...
Страница 2: ......
Страница 6: ...iv Platinous User s Manual...
Страница 22: ...Precautions in Usage 16 Platinous User s Manual...
Страница 31: ...Platinous User s Manual 25 Names and Functions of Parts...
Страница 40: ...Names and Functions of Parts 34 Platinous User s Manual...
Страница 58: ...Installation 52 Platinous User s Manual...
Страница 80: ...Operation 74 Platinous User s Manual...
Страница 106: ...Troubleshooting 100 Platinous User s Manual...
Страница 107: ...Platinous User s Manual Chapter 9 Appendix...