MAW
®
User Manual Rev. 3.0
40
Each zone has an extra sensor for validation (zones 1, 2, 3, 5, 6), these sensors are not con-
nected to any of the electronics on the unit.
PT-1000 Class-B sensors are located in the central positions of bottom zones and are con-
nected to connectors on the side of the table plate as shown below:
Figure 21.1
Pin
A connector and wire may be supplied by Esco Medical or your distributor.
External temperature conditions of zones can be constantly logged through the connectors,
without compromising any performance of the unit. Any logging system that uses standard
PT-1000 sensors may be used.
Esco Medical can supply an external logging system for the sensors.
22 All-in-one PC
The ART workstation is equipped with a powerful touch-enabled AIO PC. Switch the PC
on/off by pressing the button below the screen.
Figure 22.1
The screen