TL family`s multiroom IVF incubators User Manual Rev. 1.0
81
2.
VOC/HEPA filters must be replaced every 3 months.
3.
In-line HEPA filters must be replaced yearly during annual maintenance.
4.
According to the intervals prescribed by the clinical practice in the laboratory
where the MIRI
®
TL6 or MIRI
®
TL12 multiroom IVF incubator is in use, Suitable
cleaning is in use. The manufacturer does not recommend periods longer than 14
days between cleanings.
It is essential to perform the inspection and service at the intervals indicated
in the “38 The
Maintenance guide
”
section. Failure to do so can have a severe
adverse outcome, causing the unit to stop performing as expected and causing
damage to samples, patients or users.
Warranty is considered to be void if service and maintenance procedures are
not followed.
Warranty is considered void if service and maintenance procedures are done
not by trained and authorized personnel
.
29 Emergency procedures
Total loss of power to or inside the unit:
•
Remove all the samples and place them in an alternative or backup device that is
not affected by the problem.
•
Without the power source, the internal temperature of
the MIRI
®
TL6 and MIRI
®
TL12 multiroom IVF incubator
will drop below 35 °C after being 10 minutes in an
ambient environment of 20 °C.
•
The CO
2
concentration will remain within 1% of the setpoint for 30 minutes if the
lids remain closed.
•
If a longer time to turn the power back on is needed, it may be useful to cover the
unit with insulating blankets
to slow the temperature drop.
If a single temperature alarm goes off:
•
Remove the samples from the affected compartment. They can be relocated to any
of the other compartments, which happens to be unoccupied. All compartments
are separate so that the remaining ones will function normally.
If multiple temperature alarms go off:
•
Remove the samples from the affected compartment. They can be relocated to any
of the other compartments, which happens to be unoccupied. All compartments
are separate so that the remaining ones will function normally.