Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
47
1.
It is recommended to clean the unit with an aqueous 70% isopropyl alcohol. Mois-
ten a sterile wipe and clean all the device's internal and external surfaces by rub-
bing the surfaces' wipe.
2.
After wiping, leave the device's lids open for some time to ensure that all alcohol
fumes evaporate.
3.
Finally, use purified or sterile water is used to wipe the surfaces of the device.
4.
Inspect the device
–
if visually clean, consider it ready for use.
If the device is not visually clean, repeat the process from step 1.
21.3 Manufacturer recommended disinfection procedure
Disinfection of the device (with no embryos inside)
The use of gloves and good handling techniques are essential for successful disinfection.
Proceed with the following steps (this procedure has been demonstrated during the on-
site training program as part of the installation protocol):
1.
Power off the Mini MIRI
®
Dry or Mini MIRI
®
Humidity multiroom IVF incubator
(rear panel).
2.
Open the lids.
3.
Use the required disinfectant to disinfect the internal surface and a glass plate on
the lid's top. Use sterile wipes to apply the disinfectant.
4.
Wipe all internal surfaces and the top of the lid with three wipes at least. Repeat
until the wipes are no longer discolored.
5.
Change your gloves, and after 10 minutes of contact time, spray sterile water on
the surfaces and wipe them with a sterile wipe.
6.
Inspect the device
–
if it is visually clean, consider it ready for use. If the device is
visually not clean, go to step 3 and repeat the procedure.
7.
Turn on the Mini MIRI
®
Dry or Mini MIRI
®
Humidity multiroom IVF incubator
(rear panel).
22 Heat optimization plates.
Insert the heat optimization plate. The heat optimization plate will ensure full contact
with the dish. It generally means much more stable temperature conditions for the cells.
The plate fits the compartment. The plate fits the compartment, and it is taken out for
cleaning.
Do not autoclave at the inserts. It will damage the inserts as high temperature
bends them out of shape.