16 Changing the set points
16.1 The temperature set point
The temperature setpoint can be adjusted in the range between 24.9 °C to 40.0
°C.
The default temperature setpoint is 37.0
°C
.
To change the temperature setpoint, follow these instructions:
1.
When the display shows the current temperature:
2.
Press the rotatory button and rotate (
↻
) or (
↺
) to adjust the setpoint.
3.
After changing the temperature value, press the rotatory button to save the set-
point.
If the display does not show the current temperature reading, rotatory button rotation
(
↻
) or (
↺
) will toggle between the temperature, CO
2
, O
2
and mode readings.
16.2 The CO
2
gas concentration set point
The CO
2
concentration can be adjusted in the range between 1.9% to 9.9%.
The default CO
2
setpoint is 6.0%.
To change the CO
2
gas concentration setpoint, follow these instructions:
1.
When the display shows the CO
2
gas concentration:
2.
Press the rotatory button and rotate (
↻
) or (
↺
) to adjust the setpoint.
3.
After changing the value, press the rotatory key once more to save the value.
If the display does not show the current temperature reading, rotatory button rotation
(
↻