Alarm key which indicate the alarm condition
The audio pattern is 3 short beeps followed by a 3-second pause. All alarms have the same
audio pattern.
15.1 Temperature alarms
All 2 compartments can trigger a temperature alarm if their temperature varies over ±0.5
°C from the setpoint.
Remember that changing the setpoint more than ±0.5 °C from the current tem-
perature will result in an alarm. The same goes for all calibration adjustments.
The number will indicate the zone triggering the alarm following "A".
Temperature is too high in compartment 2:
Temperature is too low in compartment 1:
The display will lock on the alarm condition and will stop alternating between the stand-
ard status messages. If the mute key is pressed, the display will shift to normal status and
show the parameters for 5 minutes until the audio alarm comes back on again. The mute
alarm key will still show the alarm condition by blinking red while the alarm is muted.
The zone layout and sensor placement are described in the section "16 Surface tempera-
tures and measuring temperature".
If a temperature sensor malfunctions, it will be indicated by the following warning: