Instructions for Safe Use
☛
external magnetic and RF disturbance exceeding the specified limits of
the shielding elements and possible magnetic compensation were taken
into account during the pre-installation phase
☛
deterioration of shielding elements.
In this case, the user should contact Esaote-authorized service personnel
for assistance.
Peripheral equipment compatibility test
The manufacturer of the peripheral equipment can assess the influence of
the O-scan system on the relative device using the special protocol
“susceptibility test”, the procedure of which is described in the chapter
“System management” in the user interface manual.
The test is aimed at assessing susceptibility to electromagnetic
disturbance of the peripheral equipment located in the vicinity of the
O-scan system; the test does not provide guarantees regarding correct
operation of the peripheral device.
This test also enables the user to verify the susceptibility of the O-scan
system to electromagnetic disturbance caused by peripheral equipment
located in the vicinity.
The “susceptibility test” protocol has been drawn up to enable operation of
the O-scan system with a high RF transmission field and high bandwidth
and gradient response speeds. The test can be carried out as per standard
procedure on the system.
The test configuration envisages the use of:
☛
Knee Coil 1
☛
geometric phantom. The geometric phantom is filled with a water
solution of NiCl
2
and NaCl
☛
Susceptibility
protocol, containing the sequences:
•
Spin Echo 18
, to assess susceptibility of peripheral equipment to
electromagnetic disturbance caused by the system. The parameters
are: TE = 18 ms, TR = 530 ms, FOV = 100*100, acquisition matrix
= 256*256, number of slices = 14, slice thickness = 2 mm, gap =
1 mm, number of excitations = 6.
W a r n i n g
Artifacts on images can be caused by loss in efficiency of the
shielding elements and possible magnetic compensation. In this case,
contact Esaote-authorized service personnel for assistance.
Содержание O-scan eXP
Страница 1: ......
Страница 2: ......
Страница 3: ...O scan User Manual 350004900 Rev 07 Esaote S p A Genoa...
Страница 9: ...350004900 Rev 07 7 10...
Страница 10: ...8 10...
Страница 12: ...10 10...
Страница 20: ...8 8...
Страница 32: ...8 8 Chapter 2...
Страница 51: ...350004900 Rev 07 19 48 Instructions for Safe Use fig 3 1 Top view...
Страница 52: ...20 48 Chapter 3 fig 3 2 Side view...
Страница 80: ...48 48 Chapter 3...
Страница 102: ...22 22 Chapter 4...
Страница 105: ...350004900 Rev 07 3 20 Technical Description fig 5 1 Equipment components...
Страница 122: ...20 20 Chapter 5...
Страница 126: ...4 20 Chapter 6 fig 6 3 The O scan Premium and O scan eXP patient seat with armrests fig 6 4 Armrest detail...
Страница 142: ...20 20 Chapter 6...
Страница 155: ...350004900 Rev 07 7 8 Coils fig 8 5 Spherical uniform phantom fig 8 6 Cylindrical uniform phantoms...
Страница 156: ...8 8 Chapter 8 fig 8 7 Geometric phantom fig 8 8 Centering support for the spherical phantom diameter 140 mm...
Страница 160: ...4 10 Chapter 9 fig 9 4 Lumbar support cushion fig 9 5 Positioning of the lumbar support cushion on the seat...
Страница 161: ...350004900 Rev 07 5 10 Soft cushion Set fig 9 6 Use of lumbar support cushion...
Страница 184: ...10 10 Chapter 11...
Страница 232: ...4 4 Chapter 17...
Страница 236: ...4 4 Chapter 18...