i-CON 1 / 0
3BA00180 • 8.04.008 • Rev. 1
Operating instructions
6.3.4 Tip offset
(Function reserved)
6.3.5 Calibrating the i-CON 1
Use the calibration temperature (see 6.3.6 “Calibration” temperature) to exactly syn-
chronize the actual tip temperature with the displayed temperature. The following
procedure must be observed for the calibration:
■
Input of desired set temperature (Chapter 6.3. “Set temperature”).
■
Set the calibration temperature (Chapter 6.3.6 “Determination of cali-
bration temperature”).