9.8.
Display test
When ac ti vating the dis play test all seg ments of the dis play are light on. The dis play
shows “
8.8.8.8.8.8
.”
Ac ti vating the dis play test by:
• Di gi tal in put 1 (pa ra me ter 0-11)
• Di gi tal in put 2 (pa ra me ter 0-12)
9.9.
Main reset
The main re set is per for med by pres sing a key com bi na ti on at the front of the pa nel
me ter. By do ing this all pa ra me ters are set ting to the de fault va lue. The va lue of the
pa ra me ter 0-00 (in put ran ge) is not chan ging by the main re set.
Du ring the main re set the dis play shows “Init.”.
Per form the main re set by
Pres sing the push but tons
,
and
at the same time du ring 10 se conds.
10. Error codes
10.1. Encoder not connected “Err01"
• The dis play flashs and in di ca te “Err01"
• Si gna li zes that no en co der have been con nec ted to the in stru ment
10.2. Waiting for data input “Err02"
• The dis play flashs and in di ca te “Err02"
• Si gna li zes in sla ve- mo de, that af ter the con nec tion of an en co der no data in put
have been re cei ved by the in stru ment.
10.3. External clock frequency too high “ERR03"
• The dis play flashs and in di ca te “Err03"
• Si gna li zes in sla ve mode, that the clock fre quen cy of the “ex ter nal clock” is too
high (> 500 kHz).