15 Calibration
36
15
C
ALIBRATION
15.1 General provisions
Calibration mode is intended for zero and span calibration of the gas detector.
Perform calibration only in the safe area free from dangerous gases and with
oxygen content not higher than 20.9%.
Perform calibration at an ambient
temperature of 20±5 °С.
If the gas detector was used or placed
for storage at a negative temperature, hold the
gas detector within 3–4 hours at a temperature
from 0 to +30 °С before the calibration.
Perform zero calibration at a temperature and
environmental relative humidity corresponding to
the measurement conditions.
Perform span calibration at least one time per
6 months.
15.2 Calibration procedure:
Put the calibration cap to the gas detector till
it clicks as shown in the figure.