Specification and User Guide

 

Product ID:

  T1B1

 

Manual 950-104 version:

  B

Mobile 

Safe-T-Shield

Содержание Safe-T-Shield

Страница 1: ...Specification and User Guide Product ID T1B1 Manual 950 104 version B Mobile Safe T Shield...

Страница 2: ......

Страница 3: ...ialists at 1 866 ERGO 4 US We may also be reached via email at equip ergonomic products com We hope you enjoy the benefits and quality of your new equipment and look forward to handling your future ne...

Страница 4: ...system available the Safe T Shield is intended to be used in conjunction with a sensible PPE regime The requirements for such PPE use including gloves gowns and masks will vary by region and need User...

Страница 5: ...nual Doing so may result in injury or exposure to shortwave UV energy EMC WARNING As with many devices using high powered electrical motors the Safe T Shield requires special precautions with regard t...

Страница 6: ...12 A Gift for You Your Team and Your Patients Your Safe T Shield is shipped with a 90 o HVE Handpiece and a pack of disposable Short Handled Mirrors While not required for use we have found these inst...

Страница 7: ...n damage to the device and injury to the operator ASSEMBLY OVERVIEW The Safe T Shield requires only minor assembly and no tools are needed 1 Remove the filters to access and remove the packing materia...

Страница 8: ...Filter Assembly upon arrival Remove and discard this packing material prior to use fig 3 6 Failure to do so may cause damage to the motor 7 Replace filters in reverse order from which they were remov...

Страница 9: ...e as the Power Plug 3 Carefully position the lid on the top of the Housing and lower until firmly seated 4 Secure each of the four latches ATTACH POWER AND CONTROL CORDS fig 3 16 1 Connect the Power C...

Страница 10: ...ine See page 18 for cleaning and disinfection recommendations To remove the Shield 1 Depress the retaining tab on one side of the inner part of the Manifold and gently angle the Shield out until free...

Страница 11: ...ekly cleaning To remove the Airflow Plate 1 Remove the Shield from the Manifold see above 2 Depress either retaining tab holding the Plate in place fig 3 11 3 Grasp the removal ring in the center of t...

Страница 12: ...Assembly has been removed see page 8 Operating the unit with this packing material in place can cause damage to the motor If any Filters become blocked they should be replaced immediately 2 Do not blo...

Страница 13: ...s speed in increments of 1 the right side button increases it After 3 minutes without any adjustments to the vacuum speed the display on the Housing will show a 3 digit number indicating the cumulativ...

Страница 14: ...he patient chair as shown below For best results position the base no more than 24 away from the patient s head Distances greater than that may cause strain on the adjustable Suction Tube and or make...

Страница 15: ...roperly position the Shield above the patient 1 Place your open hand on the side of the Suction Tube away from you and approximately 6 above the circular Handle Place your other hand on the top of the...

Страница 16: ...nd that angled HVE handpieces provided the highest level of comfort and access to the oral cavity while allowing the Shield to best perform it s critical task of controlling the aerosols generated in...

Страница 17: ...VE tips allowed better positioning For instance assistants suctioning from the left during treatment of posterior upper left teeth Standard tips may be easily cut to any length desired as shown right...

Страница 18: ...or fogging we recommend using only the cleaning solutions and disinfection products below on the entire unit to avoid accidental contact with the Shield and Manifold Allowed Manufactured Disinfection...

Страница 19: ...y it is worth mentioning that the Airflow Plate is prominently in the patient s view during treatment To maintain a positive patient experience it is vital that this plate is not only thoroughly disin...

Страница 20: ...your manufacturer s instructions to prepare your fogger for use 4 Turn on the Safe T Shield 5 Place the fogger nozzle in the center of the Suction Tube 6 Fog the tube for 10 seconds only and stop To a...

Страница 21: ...disposal protocols when replacing filters The LED window on the Housing will display a symbol if the filter requires replacement Filter Component Frequency Primary Cotton Filter Every 40 Hours Seconda...

Страница 22: ...l blockages If no internal blockage is found replace the filters LED window displays ERR alarm sounds and the motors stops working There is a fan or internal power circuit failure Cycle the power off...

Страница 23: ...el Safe T ShieldTM item T1B1 Filtration Three part filter system Filtration Efficiency H13 Level 99 97 at 0 3 m Airflow Volume w filter 222m3 hr 130cfm UV Specification 265 185nm 2 26mW cm2 Power 110...

Страница 24: ...This page intentionally blank...

Страница 25: ...will tell it s own story We will never sell you more than our equipment can deliver Life is too short for that We promise to offer you a significant warranty on our products that we will unequivocall...

Страница 26: ...950 104revB ergonomic products 198 Airport Rd Fall River MA 02720 866 374 6487 Safe T ShieldTM Mobile Item T1B1 26 NOTES...

Страница 27: ......

Страница 28: ...or Spotlight Clients These exceptional doctors will be happy to show you the impact our high performance equipment has had on their practices With locations across the nation you aren t far from seein...

Отзывы: