Operation Manual

Version: 12/01

25

er900 PC Operation Manual

Art-Nr: 475.156

Ergometer

Type 

er900 PC

Annex A: Technical Data

Annex A:  Technical Data

Type :

ergometrics  900 PC

Operation mode:

continuous operation

Power supply:

115 V / 230 V ~/ 50 Hz / 60 Hz

Power rating:

90 VA

Braking principal:

computer controlled eddy current brakes with torque regula-
tion  (RPM independent in accordance with DIN 13405)

Load range:

20 . . . 999 Watt

RPM range:

30 . . .  130 n/min

-1

Deviation of  load :

acc. to DIN 13405: max. ± 5% in range 25 and 400 W

however:

max. ± 3 W  in range 25 and 100 Watt

acc. to manufacturer: max. ± 5% in range 20 and 900 W

however:

max. ± 3 W in range 20 and 100 W

Load increments:

manual setting: 5W programmable: 5W or multiple thereof

Max.  admissable patient weight:

120 kg

Seat adjustment:

continuously adjustable for body sizes 1.20m to 2.10m

Handle bar adjustment:

continuously adjustable for body sizes 1.20m to 2.10m

Display:

1x LCD-Display : 4 lines @ 20 char.
1 x LED 10 mm: rpm display for patient

Blood pressure measurement :

RR indirect with automatic artefact suppression

Pulse measurement :

priority principal:

1. ECG, 2. Opto, 3. RR

measurement range:

35 to  230 bpm

Electrical inputs:

Digital control: RS232

Analog:

Opto, ECG 1V

Electrical outputs:

Digital:

RS232,

Analog:

load, pulse, systole, diastole
remote start ECG (1 to 99 sec)

Weight:

64  kg

Dimensions (W x L):

min.:  400 x 900 mm, width of handle bar: approx. 540 mm
height: min: 950 mm; max: 1330 mm

Calibration:

performed with switches and keypad

Protection of calibration:

service plug

Safety standard:

BF according to IEC 601-1

Device  class:

I  according to IEC 601

Classification MDD:

class IIa according to 93/42 ECC

Содержание ER900 PC

Страница 1: ...Operation Manual ergometrics er900 PC Art Nr 475 156 Version 12 01...

Страница 2: ...is only permitted after prior written consent by the manufacturer This Operation Manual is not registered for automatic update in case of eventual alterations Information regarding the latest version...

Страница 3: ......

Страница 4: ...Version 12 01 er900 PC Operation Manual Art Nr 475 156...

Страница 5: ...Device Protecting Features 9 2 7 Measurement Accuracy Verification 9 2 7 1 Routine Control of the Blood Pressure Measurement Unit 9 2 7 2 Periodic Accuracy Verification for Blood Pressure Unit 9 2 7...

Страница 6: ...6 1 General Preparations 21 6 2 Preparations for External Control via ECG or PC 22 6 3 Start of the Ergometry Session 22 6 4 Manual Operation of Ergometry Session 22 6 5 Operational Panel for Manually...

Страница 7: ...be done at your ECG device This handbook is intended for operators of the unit It explains all the functions of the er gometer in detail and in such a manner that operation of the unit can be learned...

Страница 8: ...Operation Manual er900 PC Operation Manual 4 Version 12 01 Art Nr 475 156 Ergometer Type er900 PC...

Страница 9: ...ded via special statements by the operating com pany e g for supervision and obligatory registrations internal maintenance plans test ing intervals 1 1 Comments by the Manufacturer for the Operating C...

Страница 10: ...g Electromagnetic compatibility 1 3 Warranty Regulation ergoline GmbH is only then responsible for the safety and reliability of this device when all revisions enhancements repairs and service to the...

Страница 11: ...ergometer For such purpose special leaning ergometer or offered by ergoline c Application under outdoor conditions is not permitted d In case the ergometer is not being used according to above conditi...

Страница 12: ...sonal 2 4 Normative Notices 2 4 1 Notice for EMC electromagnetic compatibility The EMC test was done with a interference level of 3V m The device should not be used in areas with a higher level The de...

Страница 13: ...pabilities are cause for re certification 2 7 1 Routine Control of the Blood Pressure Measurement Unit The automatic blood pressure measurement unit in the console of the ergometer must be approved an...

Страница 14: ...Operation Manual er900 PC Operation Manual 10 Version 12 01 Art Nr 475 156 Ergometer Type er900 PC 2 8 Identification plate Labelling Safety Notices label with serial number...

Страница 15: ...surement head with the tin flaps into the handlebar pipe Attach grounding cable connection cable and tubing 3 Adjust black rear cover onto the measurement head and tighten with screw 4 Insert saddle p...

Страница 16: ...ric size 200 850 saddle for children incl saddle pipe and sadle bellows 475 156 er900PC Operation Manual 475 190 er900 Service Unterlagen 3 4 Transport Please ensure that the mains cable is removed pr...

Страница 17: ...ation the device Minimal fuse value 16A 4 1 Operating elements and Connections Putting into Operation 1 power connector fuse 2 main power switch 3 clamp handle for seat adjustment 4 clamp handle for h...

Страница 18: ...156 Ergometer Type er900 PC 4 2 Measurement head 4 2 1 Keyboard Putting into Operation 1 LCD Display Enable blood pressure measurement Initiate or stop blood pressure measurement Increase load level...

Страница 19: ...ect the unit s power plug to one of the white outlets on the utility cart The ergometer can then be switched on and off together with the suction device 4 3 2 Connector for Blood Pressure Cuff The cuf...

Страница 20: ...3 1 Digital RS 232 interface connector for PC or ECG 2 Analog interface for analog control and analog data transfer to ECG 3 Fiber optic connector for ear finger clip or HR input from and ECG The int...

Страница 21: ...gometer Type er900 PC 4 4 Power on Sequence Mains power switch The unit will subsequently execute a self test At the end of the self test the main menu will appear as below Putting into Operation P 00...

Страница 22: ...PC without an ECG device The load controll can be done through 5 Watt steps by using the 5W and 5W keys The NIBP can be controlled through the two corresponding keys Use of the Blood Pressure Module...

Страница 23: ...of the last blood pressure reading current heart rate in case of continuous HR signal t 00 min duration of the ergometry in min S 000 mmHg systolic value of the last blood pressure reading the mea sur...

Страница 24: ...Operation Manual er900 PC Operation Manual 20 Version 12 01 Art Nr 475 156 Ergometer Type er900 PC Operation...

Страница 25: ...at the test subject can comfortably sit upright To do this loosen the handlebar height clamp handle and adjust appropriately Next you can adjust the angle of the handlebar by loosening the handlebar a...

Страница 26: ...ed Configure your ECG or PC in accordance with the appropriate operator s instructions for ergometry Start the ergometric session with the ECG or PC 6 3 Start of the Ergometry Session Start the ergome...

Страница 27: ...d decreased in steps of Watt The pos sible load covers the range from 20 Watt to 995 Watts With the key Blutdruck I O the blood pressure measurement is enabled On or disabled Off The current status is...

Страница 28: ...damage the unit s finish and lead to defects Recommended are disinfectants like Fugaten spray lysoform or Promanum N 7 2 Cleaning the Blood Pressure Cuff The blood pressure cuff may also be cleaned w...

Страница 29: ...admissable patient weight 120 kg Seat adjustment continuously adjustable for body sizes 1 20m to 2 10m Handle bar adjustment continuously adjustable for body sizes 1 20m to 2 10m Display 1x LCD Displa...

Страница 30: ...peration Manual 26 Version 12 01 Art Nr 475 156 Ergometer Type er900 PC Annex A Technical Data RPM independent range in accordance with DIN 13405 RPM independent range er900 The Speed correlated Capac...

Отзывы: