- 19 -

↑ 

 

 

 

PC Mode
Ergometry
Manual
Settings

 

Select

 

Saddle

M

ain

 

Menu

i

nitial

 

screen

pC m

oDe

Use the softkeys on the right and left (

↑ ↓

) to position the 

bar cursor on PC MODE and confirm the selection with 
SELECT.

The display changes ‑ the ergometer is waiting for com‑
mands from the external ECG unit.

With the arrrow keys, the saddle height can be electrically 
adjusted on the ergoselect 200 (on the ergoselect 400, 
these keys adjust the height of the drive unit).

As soon as the ergometer receives commands from the 
controlling ECG unit or PC, the exercise test will start and 
the corresponding values will be displayed.

The exercise test can only be terminated with the corre‑
sponding command from the controlling ECG unit.

Note

•  All functions are locked while the ergometer is operating 

in PC mode, except for the saddle height adjustment and 
the blood pressure key .

•  To reactivate the saddle height adjustment function, press    

 and the arrow keys will again be displayed.

•  Additional blood pressure measurements can be initiated 

with 

.

 

PC Mode

d

isplay

 

during

 

exercise

 

test

current load in watts

most recent BP value (systolic/diastolic values) or cuff 
pressure during inflation

duration of exercise test (min)

heart rate at the time of the BP measurement (BPM)

pedal speed (RPM)

Содержание ergoselect 50

Страница 1: ...Manual ergoselect 50 Bicycle Ergometer Operator s Manual 201000147000 Version 03 2010 English...

Страница 2: ...d in this manual All trademarks appearing in this document are trademarks of their respective owners Their protection is acknowledged No part of this manual may be reprinted translated or reproduced w...

Страница 3: ...ecting the Power Cord 13 Connecting the ECG Cable 15 Preparing the Patient 16 Adjusting the Saddle and the Handlebar 16 Operation Control terminal P 17 Turning the System On 17 Operating Modes with Co...

Страница 4: ...4...

Страница 5: ...5 Declaration of Conformity...

Страница 6: ...6...

Страница 7: ...st read and understand the content of this manual before using or work on it Paragraphs with special symbols are of particular importance If unauthorized individuals open the control terminal damaging...

Страница 8: ...evice to other pieces of equipment Equipment Malfunction Cellular telephones may not be used in the immediate vicinity of the ergometer because they might interfere with the proper functioning of the...

Страница 9: ...l If the ergometer is used for other purposes the manufacturer cannot be held liable for personal injuries or property damage resulting from the unintended use of the equipment Biocompatibility The pa...

Страница 10: ...ybody connecting additional equipment to medical electrical equipment config ures a medical system and is therefore responsi ble that the system complies with the requirements for medical electrical s...

Страница 11: ...t of the handlebar angle 3 Castors 4 Speed display RPM for patient 5 Adjustment of the saddle height 6 Power switch green button 7 Levelling devices to adjust the ergometer to uneven floors 8 Sockets...

Страница 12: ...an be lifted at the saddle and rolled away on its castors To cover greater distances however we recommend the following method Disconnect the power cord from the wall outlet Rotate the handlebar towar...

Страница 13: ...longer wobbles In case of delicate flooring it is recommended to place a mat under the ergometer to protect the flooring from damage by the feet The ergometer can be transported by means of the 2 fron...

Страница 14: ...ide the power cord under the strain relief and screw the strain relief to the frame the plastic pin must lock into place in the boring provided Put the ergometer back on its feet and set the handle ba...

Страница 15: ...face connector remote control through PC or ECG Connecting the ECG Cable ergoselect ergometers can be connected to electrocardio graphs and PC based ECG systems of most manufacturers Different connect...

Страница 16: ...oximately 10 Set the handlebar to a position where the patient sitting upright on the saddle can reach it easily To do so open clamping lever 1 and set the handlebar to a suitable angle Adjusting sadd...

Страница 17: ...option is selected the initial screen of the selected operating mode e g Ergometry will be displayed instead of the main menu With the key you can display the main menu If an error message e g E 01 a...

Страница 18: ...load changes via the keypad SETTINGS Used to configure the ergometer Speed readout At the top of the control terminal there is a speed readout for the patient as well as three LEDs that inform the pat...

Страница 19: ...he exercise test will start and the corresponding values will be displayed The exercise test can only be terminated with the corre sponding command from the controlling ECG unit Note All functions are...

Страница 20: ...tion with SELECT The exercise test is started with the Start key a blood pressure measurement at rest may precede the test see Settings When the basic load appears on the display after approx 15 secon...

Страница 21: ...e selected manually It is recommended that the patient continues to pedal in the recovery phase The END key in the middle will terminate the test End Note The saddle height ergoselect 200 can be chang...

Страница 22: ...easurements The exercise test is started with the Start key afterwards the load can be set and changed with the 5 W and 5 W keys in increments of 1 W up to 25 W as config ured Blood pressure measureme...

Страница 23: ...the ergometer will display this menu Use the softkeys on the right and left to position the bar cursor on your preferred default mode and save the selection with SELECT Protocols Protocols 6 15 are us...

Страница 24: ...is adjustable in the range from 0 to 100 Load Change Here you determine the increments for each load change Depending on your choice each key press will change the load by 1 5 10 und 20 Watts Selecti...

Страница 25: ...in with you select DATE or TIME and confirm the selection Then the value displayed in reverse video can be edited with the keys and saved with SELECT The time is adjusted in the same way You exit the...

Страница 26: ...be initiated with digital commands Digital default The communication with the ergo meter is entirely controlled with digital commands Analog IN OUT The entire communication load control and BP mea su...

Страница 27: ...e RPM limit values RPM Min 0 70 54 Max 50 130 64 Select Note The limits selected in this menu only apply to the load range between 6 and 150 watts At higher loads the RPM limits automatically adapt to...

Страница 28: ...28...

Страница 29: ...d with soap water The cloth should only be moist and not dripping wet If the cleaning agents and disinfectants used are caustic or contain alcohol they may damage and or discolor the upholstery Only t...

Страница 30: ...o years according to the rules of the art by a Service Engineer authorized by ergoline Similarly the automatic sphygmomanometer in the control terminal must be checked and calibrated by an authorized...

Страница 31: ...ad Increments user programmable Internal Protocols Control Terminal P 5 fixed incremental exercise test protocols e g WHO 10 user programmable protocols manual load control Permitted Patient Weight 16...

Страница 32: ...on temperature 10 to 40 C 50 to 104 F rel humidity 30 to 75 no condensation atmospheric pressure 700 to 1060 hPa transport and storage temperature 40 to 70 C 40 to 158 F rel humidity 10 to 90 no conde...

Страница 33: ...black speed independent range to DIN VDE 0750 0238 black grey speed independent range of the ergoselect ergometer Family of characteristics of load duration due to IEC 60601 1 Under permanent load co...

Страница 34: ...34...

Страница 35: ...operation in the configuration in which it is being used Guidance and Manufacturer s Declaration Electromagnetic Emissions The ergoselect ergometer is intended for use in the electromagnetic environm...

Страница 36: ...differential mode 2 kV common mode 1 kV 2 kV Mains power should be that of a typical commer cial or hospital environment Voltage dips short interrup tions and voltage variations on power supply input...

Страница 37: ...from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment mark...

Страница 38: ...t Power of Transmitter W Separation Distance According to Frequency of Transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0...

Страница 39: ......

Страница 40: ...ergoline GmbH Lindenstra e 5 72475 Bitz Germany Tel 49 0 7431 98 94 0 Fax 49 0 7431 98 94 128 e mail info ergoline com http www ergoline com...