9 Control Terminal T
ergoselect 4 / 5
67
9.2.5 Manual
In this operating mode the user has complete control over
the ergometer and initiates blood pressure measurements.
Manual
Pressing the [ Start ] key initiates the exercise test, the
[ + 5 W ] and [ – 5 W ] keys are used to control the load.
The actual load change can be set in the configuration
menu b/– 1 W and +/– 25 W (see chapter 9.2.6
“Setup”, section “Load Change” on page) 69.
A blood pressure measurement is initiated with the [ RR ] key.
W
200
150
100
50
0
0
5
10
15
20
25
1/min
200
150
100
50
0
0
1/min
0
%
00:00
mm:ss
0
1/min
0
W
mmHg
--/--
Waiting for start
Start
– 5 W
+ 5 W
RR
Figure 9 – 56: Starting a manual test
Terminating the Manual Mode
The exercise test can be terminated manually at any time
with the [ Stop ] key.
The load will immediately drop to 0 watt.
There is no recovery phase in the manual mode.
W
200
150
100
50
0
0
5
10
15
20
25
1/min
200
150
100
50
0
0
1/min
0
%
00:00
mm:ss
0
1/min
0
W
mmHg
--/--
End
Stop
– 5 W
+ 5 W
Figure 9 – 57: Terminating a manual test
Содержание ergoselect 4
Страница 1: ...ergoselect 4 5 Ergometer Operator s Manual 201000433000 Version 2021 06 01 Rev 04 English...
Страница 2: ......
Страница 95: ......