ELECTROSURGERY
QUICK GUIDE
VIO
®
300 S
This quick guide is no substitute for briefing and user manual. This quick guide does not include any safety notes. Make yourself familiar
with the safety notes in the user manuals.
Necessary operating steps
Check the settings of the instrument receptacles used by pressing their
focus buttons one by one.
Only if you are using a monopolar instrument.
The long edge of the return electrode
(1)
must point to the operating
field.
Following application and during surgery.
If the red return electrode symbol lights up, check the following:
•
Is the return electrode correctly connected to the device?
•
Does the long edge of the return electrode point to the operating
field?
•
Does the return electrode have sufficient contact to the skin?
!
1
Switch on the device
2
Select and adopt program
3
Check settings
max. Watts
COAG
Effect
+
+
CUT
Effect
max. Watts
+
+
Prog/
Setup
HIGH CUT
FORCED
COAG
2
MONOPOLAR
Effect
max. Watts
max. Watts
Effect
4
Apply and connect return electrode
5
Monitor return electrode
Correct
Error
6
Connect instruments
ATTENTION:
You may only plug monopo-
lar 4 mm Ø connectors into the receptacle
with the blue ring
(1)
. Otherwise you
damage the device.
1
1