ELECTROSURGERY
QUICK GUIDE
VIO
®
200 D
This quick guide is no substitute for briefing and user manual. This quick guide does not include any safety notes. Make yourself familiar
with the safety notes in the user manuals.
Necessary operating steps
The long edge of the return electrode
(1)
must point to the operating
field.
If the red return electrode symbol lights up, check the following:
•
Is the return electrode correctly connected to the device?
•
Does the long edge of the return electrode point to the operating
field?
•
Does the return electrode have sufficient contact to the skin?
!
1
Switch on the device
2
Select program
3
Check settings
4
Apply and connect return electrode
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
5
Monitor return electrode
Correct
Error
6
Connect instruments
ATTENTION:
You may only plug monopo-
lar 4 mm Ø connectors into the receptacle
with the blue ring
(1)
. Otherwise you
damage the device.
7
For APC applications, open argon bottle
1