DISMANTLING AND ASSEMBLING

LAP BiSect scissors

Dismantling the product

The product must be cleaned and disinfected in disassembled condition with jaws opened.

1   Pull out the connecting cable  

(no illustration).

2  Hold the handle (

a

) in such a way that the 

moving handle is not obstructed. Press the 

push-button and withdraw the shaft (

c

from the handle.

3   Hold the electrode jaws (

b

) between your 

thumb and index finger. Screw the electrode 

out of the shaft by turning the rotary wheel 

(

e

) (approx. 6 turns).

4 Withdraw the electrode (

d

) from the shaft.

5 Pull the insulating tube (

b

) off the shaft.

Assembling the product

1   Push the insulating tube (

b

) onto the shaft 

(

c

).

2   Insert the electrode (

d

) into the shaft up to 

the stop position.

3  Hold the electrode jaws (

b

) between your 

thumb and index finger. Screw the electrode 

in, tightening it well, by turning the rotary 

wheel (

e

) (approx. 6 turns); do not jam the 

thread.

4   Hold the handle (

a

) just like you would hold 

a pen in such a way that the handles face 

upwards and the moving handle is not ob-

structed, and press the push-button. Insert 

the shaft into the handle (

a

) until it snaps 

into place, without jamming it.

The complete instrument

The four disassembled components:

a Handle b Insulating tube c Shaft d Electrode

c

a

d

b

c

d

b

a

For the complete notes on use, refer to 

erbe-med.com

a

c

b

d

b

e

b

e