background image

80114-381_V23323

2020-07

HYDROSURGERY

QUICK GUIDE

ERBEJET

®

 2

This quick guide is no substitute for user manuals and notes on use. This quick guide does not contain any safety instructions. Familiarize
yourself with the safety instructions in the user manuals and notes on use.

For all instruments:

Switch on the ERBEJET 2 unit.

1. Use the selection buttons 



 to select a

program for your application. 

2. Confirm using the 

OK

 selection button.

Connect the fluid tube connector and
pump cartridge. The connector must audi-
bly click into place.

Note: 

For sterile procedures, the connection

should take place in a sterile area.

1. Wait until the ERBEJET 2 display shows

Please insert new cartridge!

2. Slide the pump cartridge into the pump

shaft. The pump cartridge must audibly
click into place.

Only for HybridKnife, HybridAPC

probe and monopolar applica-

tor:

1. Switch on the VIO.
2. Select a program for your application.

Connect the return electrode to the VIO.

Insert the HF connector into a suitable
socket (see above).

Insert the FiAPC connector into the APC
socket.

For all instruments:

1. Connect an infusion set to a bottle or a

bag with separating medium. The sepa-
rating medium is sterile, physiological
saline solution.

2. Hang the bottle or the bag. 

Note:

 Hang the bottle or the bag a maximum

of 1 m above the standard rail. The lowest
possible attachment point is at the height of
the standard rail.
3. Open the roller clamp.
4. Fill the infusion set until it has been fully

primed.

5. Remove the protective cap from the pump

cartridge.

6. Connect the pump cartridge and infusion

tube.

!

1

Switching on ERBEJET 2

2

Selecting the program

3

Connecting the fluid tube 

connector and pump cartridge

4

Inserting the pump cartridge

5

Select program on VIO

6

Connecting the neutral electrode

7

Connect the HybridKnife or 

monopolar applicator to the VIO

8

Connect the HybridAPC probe to 

the APC unit

MO /

MONOPOLAR

MF-U

APC

9

Connect the separating medium to 

the pump cartridge