ULTRA-E-F
PART-3 TRANSMITTER INSTALLATION AND OPERATION
INSTRUCTIONS
Transmitter Installation
After unpacking, it is recommended to save the shipping carton and packing materials in case
the instrument is stored or re-shipped. Inspect the equipment and carton for damage. If there is
evidence of shipping damage, notify the carrier immediately.
The enclosure should be mounted in an area that is convenient for servicing, calibration or for
observation of the LCD reading.
1 Locate the transmitter within the length of transducer cable that was supplied with the
ULTRA-E-F system. If this is not possible, it is recommended that the cable be exchanged for
one that is of proper length. Transducer cables that are up to 990 feet [300 meters] may be
accommodated.
2.
Mount the ULTRA-E-F transmitter in a location that is:
♦
Where little vibration exists.
♦
Protected from falling corrosive fluids.
♦
Within ambient temperature limits -40 to 131°F [-40 to 55°C]
♦
Out of direct sunlight. Direct sunlight may increase transmitter temperature to above the
maximum limit.
3.
Mounting: Refer to
Figure 3.1
for enclosure and mounting dimension details. Ensure that
enough room is available to allow for door swing, maintenance and conduit entrances.
Secure the enclosure to a flat surface with four appropriate fasteners.
Figure 3.1 Mechanical Dimensions