PSI 10000 4U Serie
1.9.7
USB port (rear side)
The USB-B port on the back side of the device is provided for communication
with the device and for firmware updates. The included USB cable can be used
to connect the device to a PC (USB 2.0 or 3.0). The driver is delivered with the
device and installs a virtual COM port. Details for remote control can be found
on the web site of the manufacturer or on the included USB stick.
The device can be addressed via this port either using the international stan-
dard ModBus RTU protocol or by SCPI language. The device recognizes the
message protocol used automatically.
If remote control is in operation the USB port has no priority over either the
interface module (see below) or the analog interface and can, therefore, only
be used alternatively to these. However, monitoring is always available.
1.9.8
Interface module slot
This slot on the rear side of the device is can receive various modules of the
IF-AB interface series. The following options are available:
Article nr Name
Description
35400100
IF-AB-CANO
CANopen, 1x Sub-D 9pole male
35400101
IF-AB-RS232
RS 232, 1x Sub-D 9pole male (null modem)
35400103
IF-AB-PBUS
Profibus DP-V1 Slave, 1x Sub-D 9pole female
35400104 IF-AB-ETH1P
Ethernet, 1x RJ45
35400105
IF-AB-PNET1P
ProfiNET IO, 1x RJ45
35400107 IF-AB-MBUS1P ModBus TCP, 1x RJ45
35400108 IF-AB-ETH2P
Ethernet, 2x RJ45
35400109 IF-AB-MBUS2P ModBus TCP, 2x RJ45
35400110
IF-AB-PNET2P
ProfiNET IO, 2x RJ45
35400111
IF-AB-CAN
CAN 2.0 A / 2.0 B, 1x Sub-D 9-pole, male
35400112
IF-AB-ECT
EtherCAT, 1x RJ45
The modules are installed by the user and can be retrofitted without problem. A firmware update of the device may
be necessary in order to recognize and support certain modules.
If remote control is in operation the interface module has no priority over either the USB port or the analog interface
and can, therefore, only be used alternately to these. However, monitoring is always available.
Switch off device before adding or removing modules!
1.9.9
Analog interface
This 15-pole Sub-D socket on the back side of the device is provided for remote
control of the device via analog or digital signals.
If remote control is in operation this analog interface can only be used alternately
to the digital interface. However, monitoring is always available.
The input voltage range of the set values and the output voltage range of the
monitor values, as well as reference voltage level can be switched in the set
-
tings menu of the device between 0-5 V and 0-10 V, in each case for 0-100%.