background image

SimpleDist® System

Operation & Instruction Manual

©2021 Environmental Express. All rights reserved. 

July 2021  1065EE3_M Rev 2

+1.800.343.5319 or +1.843.881.6560   envexp.com

Содержание SimpleDist C6000

Страница 1: ...SimpleDist System Operation Instruction Manual 2021 Environmental Express All rights reserved July 2021 1065EE3_M Rev 2 1 800 343 5319 or 1 843 881 6560 envexp com...

Страница 2: ...k Potential Hazards and Maintenance 10 Diagrams Tubing Valve Assembly Figure 1 11 Boiling Tube Assembly Figure 2 12 Assembly Complete System Figure 3 13 System Operation Total Cyanide Distillation 14...

Страница 3: ...od will be honored from the original ship date Environmental Express will bear the cost of ground transportation both to and from the customer s location and bear the cost of any parts labor and clean...

Страница 4: ...old Kit C6210 Parts and Supplies Included Description Item 12 valve PVC Vacuum Manifold C6003B Base Supports 2 piece set C6007B 12 Flow Control Valves C6005 each Push lock design accepts 3 8 diameter...

Страница 5: ...cks Caps C6120 6pk 38 mm open top closures 3 packs Black Neoprene Washers C6120A 6pk 3 packs 2 Port Cap Inserts C6130 6pk 2 12 Piece Tubing Kits C6142 Tubing kits come with 12 lengths of tubing 3 8 OD...

Страница 6: ...d until the SimpleDist System has cooled down and the collection trap is removed from the assembly 10 The power should be kept plugged into its outlet until the unit has cooled down 11 If boil over do...

Страница 7: ...OH solution may be inserted between the manifold and the vacuum source 9 Open the C6142 Tubing Kit to find the lengths of silicone tubing with tubing on one end and barbed fitting adaptors into the ot...

Страница 8: ...ical Equipment for Measurement Control and Laboratory Use Part 1 General Requirements UL 61010 1 CAN CSA C22 2 No 61010 1 2nd Edition Issued 12 July 2004 with revisions through and including 28 Octobe...

Страница 9: ...ndled in a small area with minimal radiant heat loss Users should be aware of potential dangers from heating certain types of compounds Such hazards may include explosion or the release of toxic or fl...

Страница 10: ...rews Installation Requirements Locate the HotBlock under a fume hood with a minimum face velocity of 100fpm and allow a minimum of 2 of space on all sides The following environmental conditions should...

Страница 11: ...he temperature of the sample Sample temperature will usually be 5 15 less than the display temperature Adjusting the Temperature of Your HotBlock 1 Plug the HotBlock into an approved receptacle Turn t...

Страница 12: ...certain each power outlet is earth grounded at the grounding pin 5 See individual block specifications for power requirements 6 Application of the wrong supply voltage can create a fire hazard and a p...

Страница 13: ...ASSEMBLY DRAWING PROPRIETARY AND CONFIDENTIAL FLOW CONTROL VALVE BARBED FITTING ADAPTER SIMPLE DISTILLATION MATERIAL FINISH DO NOT SCALE DRAWING PROHIBITED COMMENTS SHEET 1 OF 1 Q A MFG APPR ENG APPR...

Страница 14: ...1 800 343 5319 or 1 843 881 6560 envexp com Environmental Express 12 SimpleDist Operation and Instruction Manual Diagrams Boiling Tube Assembly Figure 2...

Страница 15: ...1 800 343 5319 or 1 843 881 6560 envexp com Environmental Express 13 SimpleDist Operation and Instruction Manual Diagrams SimpleDist SimpleDist Assembly Complete System Figure 3...

Страница 16: ...od A few of these are Aldehydes Thiocyanates Nitrate nitrite Thiosulfates Chlorine Sulfides Some of these interferences are reduced or eliminated by the distillation process For example the nitrate ni...

Страница 17: ...y 5 12 Turn air vacuum valves to the off position Repeat for each sample position 5 13 Bring the volume of liquid in the collection trap up to the 25 mL line with DI water a graduated cylinder volumet...

Страница 18: ...e inlet tube with funnel tip and insert into the smaller port on cap 5 5 Pour the entire contents of the appropriate Snip Pour trapping solution into the collection trap The final analytical method wi...

Страница 19: ...with DI water a graduated cylinder volumetric flask or other volumetric glassware may be used for volume measurements 5 13 The distillate in the collection trap is now ready for analysis Pour content...

Страница 20: ...1 800 343 5319 or 1 843 881 6560 envexp com Environmental Express 18 SimpleDist Operation and Instruction Manual Care and Maintenance of Your SimpleDist Circuitry Diagram...

Страница 21: ...aps Open the valves on the top of the manifold to increase the vacuum to maintain the proper bubble rate 2 Check the volume of the sample you are using as well For cyanide distillation you should be u...

Страница 22: ...panel of the HotBlock by unscrewing the rubber feet b On the back of the controller locate the black wire at terminal 98 and white wire at terminal 99 c Set your volt meter on AC voltage d Touch your...

Страница 23: ...tBlock on and allow it to overshoot your set point temperature 4 Locate the solid state relay mounted to the bottom panel 5 Set your volt meter to measure AC voltage 6 Touch your red lead to terminal...

Страница 24: ...llow these steps 1 Power off your HotBlock and remove the bottom panel of your HotBlock by unscrewing the rubber feet 2 Locate and disconnect the white wire connected to terminal T1 of the relay and a...

Страница 25: ...onmental Express Controller SC945 110 220V 25A Solid State Relay SC952 Type K Thermocouple SC953 Terminal Board SC955 15 X 5 120V Silicone Rubber Heater Mat C6300 for model C6002 12 Place Block 15 X 5...