Conductivity Electrode (35419-47)

Instruction Manual

This manual describes the operation of the conductivity       
electrode, 35419-47. Be sure to read this manual carefully,        
before using the electrode.

Conformable standards

This equipment conforms to the following standards:

Authorised representative in EU and UK

Cole-Parmer GmbH

Futtererstraße 16

Wertheim 97877

Germany

Telephone: +49 9377 9203-0

Fax: +49 9377 1388

E-Mail: [email protected]

Website: www.coleparmer.co.de

COLE-PARMER INSTRUMENT COMPANY UK

9 Orion Court

Ambuscade Road

St. Neots PE19 8YX

United Kingdom

Telephone: +44 (0)1480 272279

Free Call: +44 (0)1480 272279

E-Mail: [email protected]

Website: coleparmer.co.uk

Cautions during handling

Do not allow the electrode to come in contact with any 
hard surface.

If the electrode pole plate is dry, immerse the electrode         
in pure (ion exchange) water for at least one hour, prior          
to use.

The electrode connector requires a high degree of       
insulation.Do not allow the connector to come in contact        
with water or dirty hands.

Do not rub the electrode plate with a brush or polish it           
with a polishing agent.

The water-resistant construction of pH meters can be       
used in combination with this electrode to provide water-        
resistant construction (conforming to IP-67). When     
measuring, do not immerse the cap, lead or connector in         
the sample.

Contents

RoHS: EN 

IEC 

63000

9. Monitoring and control instruments including 
industrial monitoring and control instruments

RoHS: BS 

EN 

IEC 

63000

9. Monitoring and control instruments including 
industrial monitoring and control instruments

Name

Q’ty

Electrode

1 pc

Instruction manual

1 copy

Specifications and parts description

Part name

Specifications

Preparations

Connecting to pH meter

Setting cell constant

The cell constant for this electrode is displayed on the         
electrode cap. 
Enter the cell constant (shown as 1.030 × 100  m

-1

,  in this           

example) of the electrode, by following the instructions in the         
Operation Manual for the meter.

 

Example:   

   35419-47

 

LOT

. ......

     1.030 x 100 m

-1

Electrode model

35419-47

Cell constant

100 m

-1

 (Former units: 1 cm

-1

)

Measurement range

0.1 mS/ to 10 S/m 
(Former units: 1 

S/cm to 100 mS/cm)

Usable temperature range

0

°

C to 80

°

C

Storage temperature range

0

°

C to 50

°

C

Electrode material

Titanium - platinum black

Wetted part materials

PPS, PSF and titanium

Electrode length

150 mm (incl. cap)

Max. height of electrode plate

53 mm (position from electrode tip)

Exit. diam, of wetted part

16 mm

Lead length

1 m

1. Insert the electrode connector   

into the connector port sleeve

    

on the meter, after aligning with     

the pin. Do not insert the     

connector unless it is aligned    

properly with the connector   

port.

2. Press the electrode connector   

into the connector port on the

     

meter, while turning the

   

connector to the right.

3. Slide the connector cover over    

the connector. Then, push the

    

cover in straight until it comes     

in light contact with the meter     

case.Do not turn the cover.

4. Insert the temperature connec-   

tor into the jack on the meter.      

Insert the connector firmly, until    

the O-ring on the connector    

can no longer be seen.

Electrode 
connector

Protective cap

Cap

Lead

Electrode plate

Connector cover

Temp. connector (T electrode)

C1

T1

T2

C2

C1

T1

T2

C2

C1

T1

T2

C2

C1

T1

T2

C2