Conductivity Electrode (35419-47)
Instruction Manual
This manual describes the operation of the conductivity
electrode, 35419-47. Be sure to read this manual carefully,
before using the electrode.
Conformable standards
This equipment conforms to the following standards:
Authorised representative in EU and UK
Cole-Parmer GmbH
Futtererstraße 16
Wertheim 97877
Germany
Telephone: +49 9377 9203-0
Fax: +49 9377 1388
E-Mail: [email protected]
Website: www.coleparmer.co.de
COLE-PARMER INSTRUMENT COMPANY UK
9 Orion Court
Ambuscade Road
St. Neots PE19 8YX
United Kingdom
Telephone: +44 (0)1480 272279
Free Call: +44 (0)1480 272279
E-Mail: [email protected]
Website: coleparmer.co.uk
Cautions during handling
Do not allow the electrode to come in contact with any
hard surface.
If the electrode pole plate is dry, immerse the electrode
in pure (ion exchange) water for at least one hour, prior
to use.
The electrode connector requires a high degree of
insulation.Do not allow the connector to come in contact
with water or dirty hands.
Do not rub the electrode plate with a brush or polish it
with a polishing agent.
The water-resistant construction of pH meters can be
used in combination with this electrode to provide water-
resistant construction (conforming to IP-67). When
measuring, do not immerse the cap, lead or connector in
the sample.
Contents
RoHS: EN
IEC
63000
9. Monitoring and control instruments including
industrial monitoring and control instruments
RoHS: BS
EN
IEC
63000
9. Monitoring and control instruments including
industrial monitoring and control instruments
Name
Q’ty
Electrode
1 pc
Instruction manual
1 copy
Specifications and parts description
Part name
Specifications
Preparations
Connecting to pH meter
Setting cell constant
The cell constant for this electrode is displayed on the
electrode cap.
Enter the cell constant (shown as 1.030 × 100 m
-1
, in this
example) of the electrode, by following the instructions in the
Operation Manual for the meter.
Example:
35419-47
LOT
. ......
1.030 x 100 m
-1
Electrode model
35419-47
Cell constant
100 m
-1
(Former units: 1 cm
-1
)
Measurement range
0.1 mS/ to 10 S/m
(Former units: 1
S/cm to 100 mS/cm)
Usable temperature range
0
°
C to 80
°
C
Storage temperature range
0
°
C to 50
°
C
Electrode material
Titanium - platinum black
Wetted part materials
PPS, PSF and titanium
Electrode length
150 mm (incl. cap)
Max. height of electrode plate
53 mm (position from electrode tip)
Exit. diam, of wetted part
16 mm
Lead length
1 m
1. Insert the electrode connector
into the connector port sleeve
on the meter, after aligning with
the pin. Do not insert the
connector unless it is aligned
properly with the connector
port.
2. Press the electrode connector
into the connector port on the
meter, while turning the
connector to the right.
3. Slide the connector cover over
the connector. Then, push the
cover in straight until it comes
in light contact with the meter
case.Do not turn the cover.
4. Insert the temperature connec-
tor into the jack on the meter.
Insert the connector firmly, until
the O-ring on the connector
can no longer be seen.
Electrode
connector
Protective cap
Cap
Lead
Electrode plate
Connector cover
Temp. connector (T electrode)
C1
T1
T2
C2
C1
T1
T2
C2
C1
T1
T2
C2
C1
T1
T2
C2