24
INVUE GV148 LIQUID CHEMICAL CONCENTRATION MONITOR
User Guide | Entegris, Inc.
Verify Compensation
The
Verify Compensation
section tab confirms
the chemistry coefficients are working correctly
by generating temperature data points for the
duration selected.
1. On the
Systems
tab, select the chemistry in the
Chemistry Applied TCC
menu.
2. On the
Temperature Compensation
tab, confirm
the chemistry number and brief description is
correct.
3. Set
Plot Interval
to record temperatures.
4. Click on the “Save to File” box to save, if desired.
5. Press “Start”. The chart should begin plotting
temperatures.
6. Press “Stop” to end verification.
7. If saved to file, assign a name to the {*.csv} file
and save.
NOTE: Due to a finite response time of the temperature
measurement, a rapid temperature transition will result
in a brief spike in the concentration reading
ANALOG/DIGITAL IO TAB
Tab allows user to force analog outputs to a specific
value and trim analog outputs to further refine specific
values.