9 Installation
9.1 Units without a Vacotron module
•
Remove the 6-series unit and any additional items ordered from the carton and inspect
for damage that may have occurred during shipment.
•
Place the unit on a desk or EN-Car. Ensure that there is sufficient air flow below the unit
(do not place the unit on a table-cover).
•
If necessary, adjust the adjustable foot [21], so that the unit does not wobble.
9.2 Units with a Vacotron module
•
Remove the 6-series unit and any additional items ordered from the carton and inspect
for damage that may have occurred during shipment.
•
Place the vacuum unit on a desk or EN-Car.
•
If necessary adjust the vacuum unit's adjustable foot [25], so that the vacuum unit does
not wobble.
•
Place the main unit on top of the vacuum unit.
•
If necessary adjust the main unit's adjustable foot [21], so that the main unit does not
wobble.
•
Push the vacuum hose [23] over the hose nipples on both units.
•
Carefully lift the main unit at the front and insert flat cable [17] into connector [18].
9.3 Connection to mains supply
•
Insert the mains cable into socket [20] and connect it to a wall socket. Status indicator
LED [14] indicates that the unit is connected to the mains supply.
!
CAUTION:
•
Do not place the unit in a location where the power cord could be tripped over or
pulled out during treatment.
•
Do not attempt to use the unit if it is not properly grounded. Make certain that the
unit is electrically grounded by connecting it only to a grounded electrical service
receptacle conforming to the applicable national and local electrical codes
regarding medical environments.
•
The unit will initialize. This may take a while. During initialization LED [14] blinks yellow.
•
At the end of the initialization the unit enters standby mode. LED [14] is now
continuously yellow.
•
Start the unit by pushing the standby button [19]
1
.
•
The unit starts by executing the self test, during which LED [14] blinks green.
•
At the end of the self test the unit enters the Home menu and is ready for use. LED [14]
is now continuously green.
•
When you have finished treatments push the standby button [19]
1
again to leave the
unit in standby mode.
1
When you push the standby button for more than two seconds, the unit will completely
reinitialize and enter standby mode.
Содержание 1600905
Страница 1: ...6 series User Manual ...
Страница 3: ...6 series User Manual ...
Страница 4: ......
Страница 5: ...3 1600751 4D ...
Страница 90: ...88 1600751 4D ...
Страница 92: ...90 1600751 4D 13 6 3 LP 13 6 4 CP 13 6 5 CPid ...
Страница 94: ...92 1600751 4D 13 7 4 IG50 Biphasic Envelope 13 7 5 IG100 Monophasic Envelope 13 7 6 IG100 Biphasic Envelope ...
Страница 95: ...93 1600751 4D 13 7 7 IG150 Monophasic Envelope 13 7 8 IG150 Biphasic Envelope ...