5.2 Sensor Replacement

5.2.1 CO Sensor

A CO sensor must be replaced when it can no longer be calibrated.  To replace a sensor, perform the following steps.

ƒ

 

Turn off the electrical power.  The sample air can continue to flow.

ƒ

 

Open the display panel and remove the two manifold retention screws and remove the manifold.  See 

Figure 9

.

ƒ

 

Remove the old CO sensor, which is the bottom most sensor, and replace it with a new sensor.

C

AUTION

:

 New sensors come with a shorting clip that must be removed for proper operation.  See 

Figure 8

.

ƒ

 

Replace the manifold.  Observe that the flowmeter reading is correct.

ƒ

 

Turn on the electrical power.

ƒ

 

Wait 30 minutes, then recalibrate the sensor per the instructions in Section 5.1.3.  As the timer is counting down

during calibration, a voltage is displayed on the screen beside the timing count.  This voltage must be between:

 

For Serial No.

 

Range

 

Adjust to

 

133 and above

 

.38V - .68V

 

.65V

 

132 and below

 

.59V - .68V

 

.64V

ƒ

 

Adjust to: .65

V

 if it is not between .38 – .68

V

.  Adjust this voltage during the last 30 seconds of the calibration

interval, using the bottom most potentiometer on the sensor circuit board.  See 

Figure 10

.

ƒ

 

Replace the manifold housing, and secure the display panel. An instrument without the manifold housing in place
is susceptible to RFI.

5.2.2 Oxygen Sensor

An oxygen sensor must be replaced when it can no longer be calibrated in clean air.  To replace the sensor, follow the
general steps given for the replacement of the CO sensor, but effect a replacement of the oxygen sensor, which is the
large sensor in the center position of the manifold.  After sensor installation, wait four hours before recalibrating.
When recalibrating, it is not necessary to make a voltage adjustment.

5.3 Humidifier Tube

The humidifier tube is used to assure that the CO and oxygen sensors are not subjected to extremely dry air for a long
period of time, which would decrease their useful life.  Change the humidifier tube when changing a CO sensor at the
completion of its useful life.
To change the humidifier tube:

ƒ

 

Remove the sheet metal cover from the top of the enclosure, exposing the humidifier tube.  See 

Figure 10

.

ƒ

 

Note the coiled arrangement of the tube.  Remove the old tube and replace it with the new one.

ƒ

 

Replace the sheet metal cover.

5.4 Dew Point Sensor

Unlike the CO and O

2

 sensors, the dew point sensor cannot be field calibrated.  To assure correct performance, the dew

point sensor should be calibrated biyearly.  To minimize instrument downtime, take advantage of the dew point sensor
exchange program available through 

ENMET

, in which a year old dew point sensor can be exchanged for a newly

calibrated sensor.  Call 

ENMET

 customer service personnel for details.

5.5 Flow Control Orifice

A 0.006-inch diameter orifice is used to set the flow rate and to drop the air pressure.  It is located where the air enters
the humidifier tube, see 

Figure 10

.  In well-maintained medical air systems, this orifice should not clog.  However, if

difficulty is experienced in maintaining flow rate with assured inlet pressure, remove air pressure from the equipment
and examine this orifice; replace it if necessary.

5.6 Particulate Filter Replacement

On instruments with serial numbers 

S

/

N

 332 and above, a filter is installed inside the enclosure, as shown in 

Figure 10

.

To replace this filter:

ƒ

 

Obtain a new filter

ƒ

 

Note the correct direction of flow, as denoted by the arrow on the filter body

ƒ

 

Remove the old filter

ƒ

 

Replace with the new filter

Содержание MED AIR 2000

Страница 1: ...MCN 167 09 20 00 MCN 176 03 06 98 MCN 200 09 04 98 MCN 204 01 29 99 MCN 237 09 21 00 MCN 274 05 01 02 MCN 327 02 03 05 MCN 332 04 15 05 MED AIR 2000 Operation and Maintenance Manual Andersen Medical...

Страница 2: ......

Страница 3: ...peration Condition 8 4 2 Alarm Set Points 8 4 3 Alarm Latching 8 4 4 Audio Defeat 9 4 5 Display 9 4 6 Operational Menu 9 4 7 Fault Indications 10 4 7 1 Low Flow Indication 10 4 7 2 Other Fault Indicat...

Страница 4: ...SPLAY 8 TABLE 4 FACTORY ALARM SET POINTS 8 FIGURE 5 MED AIR 2000 OPERATION MENU FLOW CHART 9 FIGURE 6 MED AIR 2000 MAINTENANCE MENU FLOW CHART 11 FIGURE 7 CONNECTION OF CALIBRATION GAS CYLINDER 13 TAB...

Страница 5: ...ected verified and properly packaged at our company and delivered to the carrier in good condition When it was picked up by the carrier at ENMET it legally became your company s property If your shipm...

Страница 6: ...ord A cord to supply 110 VAC to the equipment Not illustrated Audio Alarm A loud horn activated by certain alarm conditions Mounting Flanges Flanges with holes for mounting the enclosure to a vertical...

Страница 7: ...put See note below Audio Alarm Flow Meter Strain Relief for Line Cord Fault LED SELECT Switch Power LED OPTION Switch ALARM ACKNOWLEDGE ALARM DEFEAT Switch Visual Alarm O2 If Oxygen option is installe...

Страница 8: ...ld Housing The sample manifold the carbon monoxide and oxygen sensors are located under this small aluminum housing Not illustrated in figure 2 see Figure 8 2 4 Power Supply The power supply circuit i...

Страница 9: ...rs Upright plumb and level vertical orientation of the instrument is necessary for proper operation Mount the instrument on an appropriate vertical surface using the mounting flanges provided Avoid ar...

Страница 10: ...play The instrument may go into alarm briefly but the sensors stabilize quickly If the instrument persists in alarm acknowledge the alarm by pressing the ALARM ACKN AUDIO DEFEAT switch If alarm persis...

Страница 11: ...trip in the positions given in Table 3 Table 3 Outputs for 4 20mA for S N 599 and below Position Channel Function Range 16 CO Ground 4 mA 0 ppm 17 CO 4 to 20 mA 20 mA 100 ppm 18 DP Ground 4 mA 100 F 1...

Страница 12: ...is activated If the oxygen content of the sample air decreases below the deficiency alarm set point the associated red LED is lit the associated relay chances state and the audio alarm is activated If...

Страница 13: ...concentration is given in per cent by volume When sample flow is reduced below a limit the display switches from Flow yes to Flow no or from Flow y to Flow n 4 6 Operational Menu The operational menu...

Страница 14: ...nt 4 7 2 Other Fault Indications Other fault indications are associated with sensor zero and calibration activities and are described in the maintenance section 5 0 of this manual 4 8 Dew Point Sensor...

Страница 15: ...u is accessed with the OPTION and SELECT switches The maintenance menu diagram is shown in Figure 6 Maintenance Menu Flow Chart From the operational display press the OPTION switch twice run MAINTENAN...

Страница 16: ...which locks the correct digit in place and moves the cursor one digit to the right Continue this process until the four digit key is complete When a valid key is inserted in this manner the display is...

Страница 17: ...tandard and is best avoided The best zero gas is a cylinder of 20 9 oxygen in nitrogen with no CO present This is available in the calibration kit listed in section 6 0 Pressing the OPTION switch at t...

Страница 18: ...ph 5 1 2 above When this procedure is complete ATTACH SPAN GAS SPAN 20 is displayed to indicate that the correct span gas for this procedure is 20 ppm CO in a mixture of 20 9 oxygen in nitrogen Do Not...

Страница 19: ...rated with calibrating the CO sensor and the span voltage is close to but not greater then 0 38V an adjustment may be capable of being made to extend the life of the sensor Follow the procedure foe ca...

Страница 20: ...T switch to lock the digit in place and move the cursor one digit to the right When all four digits of the new key have been selected NEW XXXX is displayed Record the new key without it the maintenanc...

Страница 21: ...ur hours before recalibrating When recalibrating it is not necessary to make a voltage adjustment 5 3 Humidifier Tube The humidifier tube is used to assure that the CO and oxygen sensors are not subje...

Страница 22: ...ove Shorting Clip Figure 9 Manifold Figure 10 Location of Parts for Replacement Manifold Cover Humidifier Tube Humidifier Tube Flow Control Orifice Oxygen Sensor CO Voltage adjustment Potentiometer CO...

Страница 23: ...1000 Fuse 1 0 Amp 5x20mm 67016 1106 Sensor Oxygen 67020 1200 Sensor CO for units S N 299 and below 67016 1204 Sensor CO for units S N 300 and above 73070 009 Orifice 73108 002 Humidifier Tube 03401 00...

Страница 24: ...ET shall not be liable for any loss or damage caused by the improper use of the product The purchaser indemnifies and saves harmless the company with respect to any loss or damages that may arise thro...

Страница 25: ...te the assessment of the degree of danger that he or she is exposed to Based upon knowledge of the effects of CO the Occupational Safety and Health Authority OSHA has set limits on exposure to CO in t...

Страница 26: ...ENMET Corporation MED AIR 2000 22 Notes...

Страница 27: ...processing delays ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour All instruments sent to ENMET are subject to a minimum 3...

Страница 28: ......

Страница 29: ...___________ Your Phone _______________________ Your PO Reference Number _______________ Your FAX _______________________ Payment Terms T COD Check one T VISA MasterCard______________________ ________...

Страница 30: ......

Отзывы: