ENMET
Corporation
M
ED
A
IR
2000
12
5.1.1 Key
Entrance to the maintenance menu is guarded with a four-digit key. The factory default setting of the key is 1270.
When a valid numerical key is inserted, the user is allowed to enter the maintenance menu.
In the "
run MAINTENANCE MODE
" position
Press the
SELECT
switch; "
KEY = 0000
" is displayed.
In the "
KEY = 0000
" position, the underline cursor is under the left digit.
Press the
OPTION
switch to change the left digit; select the correct digit.
Press the
SELECT
switch, which locks the correct digit in place and moves the cursor one digit to the right.
Continue this process until the four-digit key is complete. When a valid key is inserted in this manner, the display is
transferred to the "
run AUTOMATIC ZEROING
" portion of the menu. If an invalid key is inserted, "
INVALID
" is
displayed briefly; then the
M
ED
A
IR
2000
returns to the operational display.
Example
: Key Display and Flow Chart
KEY = 1270
CO
PPM
O
2
%
O
O
S
S
KEY=0000
Invalid
From Operational Menu
run
MAINTENANCE
MODE
Changes digit indicated by underscore cursor
Locks underscore digit and moves cursor to next digit
To Operational
Display
Valid entry
To Automatic Zeroing
O
= Option Switch
S
= Select Switch
DEW
POINT
Содержание MED AIR 2000