4.4 Audio Defeat
With the alarms in the non-latching configuration, pressing the
ALARM ACKN/AUDIO DEFEAT
switch during an
alarm silences the audio alarm.
With an alarm in the latching configuration, pressing the
ALARM ACKN/AUDIO DEFEAT
switch during an alarm
silences the audio alarm and unlatches the associated relay(s).
4.5 Display
In clean air with the correct dew point, the display is as shown in
Figure 4
, for the sensor(s) installed in the
M
ED
A
IR
2000
.
This position of the display is termed the "
operational display
". As explained below, the display can be
changed to furnish other information by using the
OPTION
and
SELECT
switches.
Concentrations of CO are given in PPM (parts per million parts of air). Dew point is given in degrees Fahrenheit at 55
PSIG
; this can be changed to degrees Centigrade by pressing the
SELECT
switch. Oxygen concentration is given in
per cent by volume. When sample flow is reduced below a limit, the display switches from "Flow: yes" to" Flow: no",
or from "Flow: y" to "Flow: n".
4.6 Operational Menu
The operational menu allows the user to:
View alarm set point concentration values
View alarm latching configurations
Enter the maintenance menu with the proper key.
The operational menu is accessed with the
OPTION
and
SELECT
switches. The operational menu flow chart is
shown in
Figure 5
,
The
OPTION
switch is indicated with a "
O
"
The select switch is indicated with a "
S
".
If the instrument is left at any location in the operational or maintenance menus, other than the operational display,
with no action taken for a period of 45 seconds, it returns to the operational display.
Figure 5: M
ED
A
IR
2000
Operation Menu Flow Chart
O
O
S
S
Operational Display
DEW POINT
°
C
↔
°
F
000 PPM
20.9%
037
°
F
FLOW:y
View
Alarms & Latches
O
S
run Maintenance
Menu
To Key
O
S
ppm CO
:
10
Latch
:
Yes
No Function
O
°
F DP
:
39.0
Latch
:
Yes
S
No Function
S
O
%O
2
:
19.5 23.5
Latch
:
Yes Yes
No Function
O
= Option Switch
S
= Select Switch
Содержание MED AIR 2000
Страница 2: ......
Страница 26: ...ENMET Corporation MED AIR 2000 22 Notes...
Страница 28: ......
Страница 30: ......