3.3.1 Non-Latching Relays
There is one non-latching, single pole relay per channel. These relays furnish normally open, normally closed, and
common terminals, which are available on TB3 and TB4; refer to the motherboard drawing. These relays are fail-safe;
the power off position and the alarm position of the contacts are identical. Contact positions are given for the power
off, alarm condition.
Table 3: Non-Latching Relay Wiring
Non-powered, alarm positions are given
Channel
Terminal Position
Contact
1
TB3 – 1NC
Normally closed
1
TB3 – 1NO
Normally open
1
TB3 – 1COM
Common
2
TB3 – 2NC
Normally closed
2
TB3 – 2NO
Normally open
2
TB3 – 2COM
Common
3
TB4 – 3NC
Normally closed
3
TB4 – 3NO
Normally open
3
TB4 – 3COM
Common
4
TB4 – 4NC
Normally closed
4
TB4 – 4NO
Normally open
4
TB4 – 4COM
Common
3.3.2 Latching Relays
Latching relays are available on TB7, the terminal strip located on the auxiliary relay circuit assembly located on the
upper surface of the control unit enclosure. These relays, once triggered by an alarm condition, remain latched in the
alarm position until released by an operator, who activates the relay release pushbutton switch on the front of the
enclosure. This switch releases all latched relays together. These relays are not failsafe; the power off position of the
contacts is opposite from the latched alarm position. The contact positions given are for the power off, non-alarm
condition
Table 4: Latching Relay Wiring
Contacts are in non-powered, non-alarm position
Channel
Terminal Position
Contact
1
TB7 – 1
Normally closed
1
TB7 – 2
Normally open
1
TB7 – 3
Common
2
TB7 – 4
Normally closed
2
TB7 – 5
Normally open
2
TB7 – 6
Common
3
TB7 – 7
Normally closed
3
TB7 – 8
Normally open
3
TB7 – 9
Common
4
TB7 – 10
Normally closed
4
TB7 – 11
Normally open
4
TB7 – 12
Common