Waterpilot FMX21
Description of device parameters
Hauser
107
Factory setting
22 mA
Set min. current
Write permission
Operators/Service engineers/Expert
Description
Enter lower current limit.
Some switching units accept no current smaller than 4.0 mA.
Options
• 3.8 mA
• 4.0 mA
Factory setting
3.8 mA
Get LRV
Write permission
Operators/Service engineers/Expert
Description
Set the lower-range value – reference pressure is present at the device. The pressure for
the lower current value (4 mA) is present at the device. Use the "Confirm" option to assign
the lower current value to the applied pressure value.
Prerequisite:
Pressure measuring mode
Options
• Cancel
• Confirm
Factory setting
Cancel
Set LRV
Write permission
Operators/Service engineers/Expert
Description
Set the pressure value, level or content for the lower current value (4 mA).
Factory setting
• 0.0 % in Level measuring mode
• 0.0 mbar/bar or in accordance with ordering information in Pressure measuring mode
Get URV (pressure measuring mode)
Write permission
Operators/Service engineers/Expert
Содержание Waterpilot FMX21
Страница 143: ......
Страница 144: ...www addresses endress com 71593280 71593280...