Waterpilot FMX21
Description of device parameters
Hauser
99
Prerequisite
• "Level selection" = In pressure
• "Calibration mode" = Dry -> entry
• "Calibration mode" = Wet -> display
Factory setting
URL of the sensor module
Full height
Write permission
Operator/Maintenance/Expert
Description
Enter the height value for the upper calibration point (vessel full). The unit is selected via
the "Height unit" parameter.
Prerequisite
• "Level selection" = In height
• "Calibration mode" = Dry -> entry
• "Calibration mode" = Wet -> display
Factory setting
Upper range limit (URL) is converted to a level unit
Density unit
Write permission
Maintenance/Expert
Description
The measured pressure is converted to a height using the "Height unit", "Adjust density" and
"Process density" parameters.
Factory setting
g/cm
3
Adjust density
Write permission
Operator/Maintenance/Expert
Description
Enter the density of the medium used to perform the calibration. The measured pressure is
converted to a height using the "Height unit" and "Adjust density" parameters.
Input: Auto dens. corr. = Off
Display: Auto dens. corr. ≠ Off
Factory setting
1.0
Process density
Write permission
Operator/Maintenance/Expert
Содержание Waterpilot FMX21
Страница 143: ......
Страница 144: ...www addresses endress com 71593280 71593280...