Waterpilot FMX21
Description of device parameters
Hauser
93
Press. eng. unit
Write permission
Operators/Service engineers/Expert
Description
Select the pressure engineering unit. If a new pressure engineering unit is selected, all
pressure-specific parameters are converted and displayed with the new unit.
Options
• mbar, bar
• mmH2O, mH2O, inH2O
• ftH2O
• Pa, kPa, MPa
• psi
• mmHg, inHg
• kgf/cm
2
Factory setting
mbar or bar depending on the nominal measuring range of the sensor module, or as per
order specifications
Temp. eng. unit
Write permission
Service engineers/Expert
Description
Select the unit for the temperature measured values.
Options
• °C
• °F
• K
Note
The setting affects the unit for the "Sensor temp." parameter.
Factory setting
°C
Sensor temp.
Write permission
No write permissions. Parameter is read only.
Description
Displays the temperature currently measured in the sensor module. This can deviate from
the process temperature.