Waterpilot FMX21
Commissioning
Hauser
67
Menu Setup → Level (LEVEL)
• RIA15 parameter: LEVEL
• Corresponds to device parameter: Level before linearization
• Visible with "Level" option, MODE = HART, device is connected
• Description:
This menu contains the parameters for configuring the pressure measuring device for
hydrostatic level measurement.
The basic settings for the device can be made via the RIA15 with this menu.
Once the LEVEL menu item is opened, the following parameters are automatically
adjusted in the device for easier operation:
• Measuring mode: Level
• Calibration mode: Dry
• Level selection: In pressure
• Lin mode: Linear
It is possible to reset these parameters to the factory default settings by performing a
reset.
Menu Setup → Level (LEVEL) → PUNIT
• RIA15 parameter: PUNIT
• Corresponds to device parameter: Press. eng. unit
• Values (default in bold)
•
mbar
•
bar
• kPa
• PSI
• Description: Use this function to select the unit for pressure
Menu Setup → Level (LEVEL) → LUNIT
• RIA15 parameter: LUNIT
• Corresponds to device parameter: Output unit
• Values (default in bold)
•
%
• m
• inch
• feet
• Description: Use this function to select the unit for level
Menu Setup → Level (LEVEL) → TUNIT
• RIA15 parameter: TUNIT
• Corresponds to device parameter: Temperature unit
• Values (default in bold)
•
°C
• °F
• K
• Description: Use this function to select the unit for temperature
3)
If the measured value that is read out is too large, it is displayed as "9999.9", for example. To display a valid measured value, the pressure unit
(PUNIT) (or level unit (LUNIT)) must be set to suit the measuring range.
4)
Default: depends on the sensor nominal range or as per order specifications