Operation options
Waterpilot FMX21
38
Hauser
The "Operator code" parameter is used to lock/unlock the device.
Operator code
Navigation
Setup → Extended setup → Operator code
Read permission
Operator/Maintenance/Expert
Write permission
Operator/Maintenance/Expert
Description
Use this function to enter a code to lock or unlock operation.
User entry
• To lock: Enter a number
¹
the release code (value range: 1 to 65535).
• To unlock: Enter the release code.
Factory setting
0
Note
The release code is "0" in the order configuration. Another release code can be defined in
the "Code definition" parameter. If the user has forgotten the release code, the release code
can be made visible by entering the number "5864".
The release code is defined in the "Code definition" parameter.
Code definition
Navigation
Setup → Extended setup → Code definition
Read permission
Operator/Maintenance/Expert
Write permission
Operator/Maintenance/Expert
Description
Use this function to enter a release code with which the device can be unlocked.
User entry
A number from 0 to 9999
Factory setting
0
Note
The device setup can also be disabled on the RIA15 via a 4-digit user code.
Additional information is available in the RIA15 Operating Instructions BA01170K.
Содержание Waterpilot FMX21
Страница 143: ......
Страница 144: ...www addresses endress com 71593280 71593280...