Turbimax CUS71D
Installation
Hauser
11
5
Installation
5.1
Installation conditions
5.1.1
Installation instructions
Basin configuration
A
B
C
D
E
6°
F
A0031574
1
Basin configuration
A
Sensor
B
50 cm (1.64 ft) Minimum distance between sensor and basin rim
C
Fixed reference point, e.g. surface of water, basin rim, bridge/walkway etc.
D
Sensor offset
E
Basin depth
F
Opening angle of ultrasonic cone 6°
Installation instructions
Find a suitable installation position for the sensor in the basin. Take the following points into
consideration when selecting the installation position:
1. Ensure the distance from the basin rim is at least 50 cm (1.64 ft) (the sensor emits
ultrasonic waves in a conical pattern).
There should not be any conduits or basin wall protrusions in the measuring range
below the sensor. Scraper units that are only temporarily in this area are permitted.