Operation
Turbimax CUS52D
32
Hauser
The following filter settings are available:
Measured value filter
Description
Weak
Low filtering, high sensitivity, fast response to changes (2 seconds)
Normal (default)
Medium filtering, 10-second response time
Strong
Strong filtering, low sensitivity, slow response to changes (25 seconds)
Specialist
This menu is designed for the Hauser Service Department.
8.1.5
Solid state reference
The solid state reference can be used to check the functional integrity of the sensor.
During factory calibration, each Calkit solid state reference is matched specifically to a
special CUS52D sensor and can be used only with this sensor. Therefore, the Calkit solid
state reference and the CUS52D sensor are permanently assigned (married) to one
another.
The following Calkit solid state references are available:
• 5 FNU (NTU)
• 20 FNU (NTU)
• 50 FNU (NTU)
The reference value indicated on the Calkit solid state reference is reproduced with an
accuracy of ± 10% when the sensor is operating correctly.
The CUY52 solid state reference with approx. 4.0 FNU/NTU is used to check the function
of any CUS52D sensors. The standard is not assigned to a specific sensor and delivers
measured values in the range of 4.0 FNU ± 1.5 FNU/NTU with all CUS52D sensors.
A0035755
35
Solid state reference